AIR SAMPLE ANALYSIS REPORT
PHASE CONTRAST MICROSCOPY
CLIENT: WARREN & PANZER ENGINEERS, P.O. lUKNAKUUND IIME 24HRS
WKP FILE# 601-01 SAMPLE DATE 4«fi!002
WKP LOG l.D. # N020436 RECEIVE DATE 4/8/2002
Contract#: WTCAR2 ANALYSIS DATE 4/e/2WZ
REPORT DATE 4/7/2002
CLIENT JOB: NYC Department of Environmental Protection
CLIENT JOB #: 1079.01.01
LOCATION WTC Sampling
AIRBORNE FIBER COUNTING, PHASE CONTRAST MICROSCOPY. NIOSH 7400. NIOSH PAT Lab
ANALYSIS:
NO. 07186; NY STATE ELAP Lab NO. 11251.
collection DEVICE’ Cassette, Collection Area “ 385 sq mm;
Field Area » 0.00785 st) mm.
SAMPLE WERE COLLECTED BY; Festy Osemwegie
SUMMARY OF ANALYSIS
CLIENT VOLUME FIBERS/ FIBERS/
SAMPLE# Site# SAMPLE LOCATION (UTERS) sq mm CC
1079.01 01 001 Field Blank 0.000
002 Field Blank 0.000
{KJ3 29 Trinity Place and Thames Street 87S 0.QQ0 <0.004
004 21 Fulton and Church Street 675 0.000 <0.004
006 20A VsBsy Street and West Broadway 876 0.000 <0.004
006 16 Park place and Church Street 675 0.000 <0.004
007 12 Murray Street and Church Street 675 0.000 <0.004
008 15 Park Place and Greenwich Street 675 0.000 <0.004
OOS 11 Murray 3troet and Greenwich SlrtMit 070 0.000 <0.004
NOTES: 1.5.5 fibars/lOO fields was used to calculate the limit of detection (LOD). Results below the LOD have no statistical validity.
2. Samples were not collected by WKP Laboratories, Inc.
3. The Federal EPA Clearance Critical following an asbestos abatement project is 0.010 fibers/cc.
4. PCM air samples will be disposed of after 3 months unless otherwise directed by client in writing.
Chris Pepino Louis Lombardi
LABORATORY ANALYST LABORATORY DIRFCTOR