AIR SAMPLE ANALYSIS REPORT
PHASE CONTRAST MICROSCOPY
CLIENT: WARREN & PANZER ENGINEERS, P.O. TURNAROUND TIME 24HRS
WKP FILE# 501-01 SAMPLE DATE 4/8/02
WKP LOG I.D. # N020446 RECEIVED DATE 4/8/02
Contract #: WTCAR2 ANALYSIS DATE 4/8/02
REPORT DATE 4/9/02
CLIENT JOB: NYC Department of Environmental Protection
CLIENT JOB#: 1079.01.01
LOCATION WTC Sampling
ANALYSIS: AIRBORNE FIBER COUNTING, PHASE CONTRAST MICROSCOPY, NIOSH 7400. NIOSH PAT Lab
NO. 07186: NY STATE ELAP Lab NO. 11251.
COLLECTION DEVICE: 25 mm MCE Filter Cassette; Collection Area = 385 sq mm;
Field Area = 0.00785 sq mm.
SAMPLE WERE COLLECTED BY: Festy Osemwegie
SUMMARY OF ANALYSIS
CLIENT VOLUME FIBERS/ FIBERS/
SAMPLE# Site# SAMPLE LOCATION (LITERS) sq mm cc
1079.01.01 001 Field Blank 0.0
002 Field Blank 0.0
003 29 Trinity Place and Thames Street 675 5.1 <0.004
004 21 Fulton and Church Street 675 5.1 <0.004
005 20A Vesey Street and West Broadway 675 11.5 0.007
006 16 Park place and Church Street 675 10.8 0.006
007 12 Murray Street and Church Street 675 14.0 0.008
008 15 Park Place and Greenwich Street 675 10.2 0.006
009 11 Murray Street and Greenwich Street 675 12.7 0.007
NOTES: 1.5.5 fibers/100 fields was used to calculate the limit of detection (LOD). Results below the LOD have no statistical validity.
2. Samples were not collected by WKP Laboratories, Inc.
3. The Federal EPA Clearance Critical following an asbestos abatement project is 0.010 fibers/cc.
4. PCM air samples will be disposed of after 3 months unless otherwise directed by client in writing.