NYC 9/11 Public Portal Document
WKP
LABORATORIES, INC
AIR SAMPLE ANALYSIS REPORT
PHASE CONTRAST MICROSCOPY
CLIENT: WARREN & PANZER ENGINEERS, P.C. turnaround TIME 24HRS
WKP RLE# a)i-oi
SAMPLE DATE 4/10/02
WKPLOOI.D.# N0204S9
RECEIVED DATE 4/10/02
Contract#; WTCAR2
ANALYSIS DATE 4/11/02
REPORT DATE 4/11/02
CLIENT JOB; NYC Department of Environmental Protection
CLIENT JOB#. 1078.01.01
LOCATION S.W. Area/Site Map
ANALYSIS; AIRBORNE FIBER COUNTING. PHASE CONTRAST MICROSCOPY, NIOSH 7400. NIOSH PAT Lab
NO. 07166; NY STATE ELAP Lab NO. 11281.
COLLECTION DEVICE: Cassette; Collection Area = 366 sq mm:
Reid Area “ 0.00786 sq mm.
SAMPLE WERE COLLECTED BY; Lawrence Baer
SUMMARY OF ANALYSIS
CLIENT VOLUME FIBERS/ FIBERS/
SAMPLE# Site# ____ SAMPLE LOCATION (LITERS) sq mm CO
1079.01.01 001 Flew Blank 0.0
002 Reid Blank 0.0
003 27 Gateway Plaza Park Area 675 3.8 <0,004
004 31 South End Ave. and Albany Street 676 9.6 0.008
DOS 32 West Street and Abany Place 678 6.1 <0.004
006 33A Rector Place by the Park Place 678 3.8 <0.004
007 34 W. Thames Str^ - Mong Watarfront 673 7.0 0.004
008 37 West of Third Place 675 3.2 <0.004
NOTES: 1.6-5 fibera/l 00 ReWs mm used to calculate the limit of detection (LOD). Results below the LOD have no siafsKcal validity.
2. Samples were not collected by WKP Laboratories. Ino.
3. The Federal EPA Clearance Critical following an asbestos abatement project is 0.010 fibers/cc.
4. PCM air samples win be disposed of after 3 months unless otherwise directed by dient in writing.
DM
LABORATORY ANALYST laboratory DIRECTOR
NYC-WTC_000161733