NYC 9/11 Public Portal Document
Part V - Quality Assurance Project Plan
133-135 Greenwich Slreet/21-23 Thames Street
Airtek Project 05-0701 - September 29, 2005
utilized with this sampling approach are as follows:
• Filters S \ 10 inch quartz fiber filters such as Whatman QMA filters or their
equivalent.
• High Volume Sampler - The sampler consists of a protective housing which
contains an electric, high-speed, high-volume blower; a filter holder capable of
holding 8x10 inch filters; a flow controller capable for controlling the air-flow
rate through the instrument at 1.0 to 1.8 m3; and a flow indicating device that
works by measuring the pressure drop across the sampler such as a magnchclic
gauge, a manometer, or rotameter. The sampler should have a gabled roof design.
The air inlet should be uniform on all sides and be designed such that the effective
particle capture velocity is 20 - 35 cm/sec.
• Thermometer - A thermometer is used to indicate the temperature at the flow
rate measurement orifice when temperature corrections are used.
• Barometer - A barometer is used to indicate the barometric pressure at the flow
rate orifice when pressure corrections are used.
• Manometer - A water or oil manometer is used to calibrate the flows of the high
volume samplers.
• Timing/Control Device - A timing device capable of starting and stopping the
sampler to obtain an elapsed run time of 24 hr ±1 hr. The accuracy of the timing
device should be ±30 minutes. • Orifice Transfer Standard - The flow rate
transfer standard is a calibrated orifice traceable to NIST standards that consists of
either a set of restrictor plates or a variable restrictor that is used to deliver a series
of discrete flow rates into the inlet of the high volume sampler so that the sampler
flow indicating device on the sampler can be accurately calibrated.
10.2.2 Mercury (Gas)
Real-time monitoring for mercury will be performed utilizing an OhioLumex RA 915+
direct read instrument. The OhioLumex RA 915+ is a factory-calibrated instrument based
on a cold vapor Zeeman atomic absorption analytical process that reduces interference
from molecular absorption bands and scattered spectra. A mercury electrodeless
discharge lamp (EDL) is used as the radiation source. The sampling system consists of a
sample pump operating at 20 L/min. and a multi-path sample cell with an effective path
length of 10 meters. Equipment and supplies utilized with this sampling approach are as
follows:
• OhioLumex RA-915+ Mercury Analyzer - A portable, self-contained field
mercury analyzer used to make real time mercury vapor determinations.
• Air Intake Hose with Pre-Filter - The sampling fine and first pass filter used to
remove particulate from the sample gas.
• Absorption and Dust Filters - Spare filters used as replacements for the internal
particulate and mercury absorbing filters in the instrument.
10.2.3 Asbestos