Ambient Air Monitoring Program Section No. 10
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 54 of 116
Equipment and supplies utilized with this sampling approach are as follows:
• OhioLumex RA-915+ Mercury Analyzer - A portable, self-contained field mercury analyzer
used to make real time mercury vapor determinations.
Air Intake Hose with Pre-Filter - The sampling line and first pass filter used to remove
particulate from the sample gas.
• Absorption and Dust Filters - Spare filters used as replacements for the internal particulate
and mercury absorbing filters in the instrument.
10.2.3 Asbestos
Asbestos sample collection will be performed in accordance with NIOSH Method 7402 -
Asbestos by TEM following the recommendations for personal sampling. Samples will be
collected at a rate of six liters per minute (6 1pm) for a minimum of eight (8) hours for a resulting
total volume of approximately 2.88 m^. To ensure optimum accuracy, the sampling rate will be
adjusted, as needed, to produce a fiber density of 100 to 1,300 fibers/mm^.
Pump flow rates of each personal sampling pump, with a representative sampler in line, will be
measured before and after sample collection. The pump will be run for 10 minutes prior to
checking the flow rate. A suitable rotameter (or bubble meter) will be connected to the filter
inlet, the flow screw adjusted to the desired sampling value, and the flow rate observed on the
rotameter (or bubble meter) will be verified as constant for a minimum of 20 seconds. Elow
rates will be recorded on Eield Sampling Data Sheets or an equivalent field logbook.
Immediately prior to collecting a sample, the sampler will be fastened to a location (i.e., on a
tripod) near the individual’s breathing zone, the top cover of the filter cassette cowl extension
(“open face”) will be removed and positioned “face down” in the sampler. The joint between the
extender and monitor body will be wrapped with tape to hold the cassette securely in place and
to provide a surface to identify the cassette. By inserting a pin through the hole near the display
screen, the elapsed time indicator is reset prior to sampling.
At the end of the sampling period, the top cover and small end caps are replaced. Samples are
transported upright to the laboratory in a rigid container with packing material to prevent jostling
or damage. To avoid electrostatic forces from causing fiber loss from the sampling filter,
untreated polystyrene foam will not be used for any of the shipping container materials.
Equipment and supplies utilized with this sampling approach are as follows:
• SKC Leland Legacy Personal Sampling Pump (or equivalent) - 5 to 15 1pm
L2005-346 Rev. 1
CITY_DEP_WORBY_CANNATA_SUBP_000114695
NYC-WTC_000428378