Ambient Air Monitoring Program Section No. 10
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 53 of 116
each sampling event. The TSP quartz filters will be analyzed for ten (10) metals: antimony,
barium, beryllium, cadmium, chromium, copper, lead, manganese, nickel, and zinc.
Equipment and supplies utilized with this sampling approach are as follows:
Eilters - Preweighed, prenumbered 8 x 10 inch quartz fiber filters such as Whatman QMA
filters or their equivalent.
• High Volume Sampler - The sampler consists of a protective housing which contains an
electric, high-speed, high-volume blower; a filter holder capable of holding 8 x 10 inch
filters; a flow controller capable for controlling the air-flow rate through the instrument at 1.0
to 1.8 m3; and a flow indicating device that works by measuring the pressure drop across the
sampler such as a magnehelic gauge, a manometer, or rotameter. The sampler should have a
gabled roof design. The air inlet should be uniform on all sides and be designed such that the
effective particle capture velocity is 20 - 35 cm/sec.
Thermometer - A thermometer is used to indicate the temperature at the flow rate
measurement orifice when temperature corrections are used.
Barometer - A barometer is used to indicate the barometric pressure at the flow rate orifice
when pressure corrections are used.
Manometer - A water or oil manometer is used to calibrate the flows of the high volume
samplers.
• Timing/Control Device - A timing device capable of starting and stopping the sampler to
obtain an elapsed run time of 24 hr ±1 hr. The accuracy of the timing device should be ±30
minutes.
Orifice Transfer Standard - The flow rate transfer standard is a calibrated orifice traceable to
NIST standards that consists of either a set of restrictor plates or a variable restrictor that is
used to deliver a series of discrete flow rates into the inlet of the high volume sampler so that
the sampler flow indicating device on the sampler can be accurately calibrated.
10.2.2 Mercury (Gas)
Real-time monitoring for mercury will be performed utilizing an OhioLumex RA 915-1- direct
read instrument. The OhioLumex RA 915-1- is a factory-calibrated instrument based on a cold
vapor Zeeman atomic absorption analytical process that reduces interference from molecular
absorption bands and scattered spectra. A mercury electrodeless discharge lamp (EDL) is used as
the radiation source. The sampling system consists of a sample pump operating at 20 L/min. and
a multi-path sample cell with an effective path length of 10 meters.
L2005-346 Rev. 1
C ITY_D E P_WO RB Y_C AN N ATA_SU B P_000114694
NYC-WTC_000428377