Ambient Air Monitoring Program Section No. 10
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 55 of 116
• Battery Pack - Rechargeable lithium ion battery pack (or equivalent), fitted with a belt clip,
compatible with the personal sampling pump.
Filters - Mixed cellulose ester membrane filters, 0.45-pm pore size, with a non-conductive
cowl, supported by a cassette filter holder, suitable for connection with the personal
monitoring pump system.
• High Flow Rotometer - 2.0 to 20.0 1pm
10.2.4 Respirable Crystalline Silica and Dust
Respirable dust and crystalline sampling will be performed according to NIOSH Method 0600 -
Particulates Not Otherwise Regulated, Respirable following the recommendations for personal
sampling. Samples will be collected at a rate of approximately two liters per minute (2 1pm) for
a minimum of eight (8) hours for a resulting total air volume of approximately 1.0 m^. To ensure
optimum accuracy, the sampling rate will be adjusted, as needed, not to exceed 2 mg dust
loading on the filter.
Filters are pre-weighed to a constant weight in the controlled weighing area. If static electricity
is evident (e.g., filter not releasing easily from the forceps or it attracts the balance pan), the filter
should be passed over an anti-static radiation source. Filters are assembled in the filter cassettes
and the cassettes closed firmly by placing a plug at each opening of the cassette to prevent
leakage around the filter. For at least 2 hours prior to sampling, the filters are equilibrated in
either an environmentally controlled weighing area or chamber.
Pump flow rates of each personal sampling pump, with a representative sampler in line and
cyclone grit pot in place, will be measured before and after sample collection. The pump will be
run for 10 minutes prior to checking the flow rate. A suitable rotameter (or bubble meter) will be
connected to the filter inlet, the flow screw adjusted to the desired sampling value, and the flow
rate observed on the rotameter (or bubble meter) will be verified as constant for a minimum of
20 seconds. Flow rates will be recorded on Field Sampling Data Sheets or an equivalent field
logbook.
The cyclone’s grit cap is removed prior to use and the interior inspected. If the inside is visibly
scored, replace the cyclone because the dust separation characteristics may be altered. Clean the
interior of the cyclone to prevent re-entrainment of large particles.
Assemble the sampler head checking for alignment of the filter holder and cyclone in the
sampling head to prevent leakage. Care must be taken to prevent the sampler assembly from
inverting at any time.
Immediately prior to collecting a sample, the sampler will be fastened to a location (i.e., on a
tripod) near the individual’s breathing zone, the top cover of the filter cassette cowl extension
(“open face”) will be removed and positioned “face down” in the sampler. The joint between the
L2005-346 Rev. 1
CITY_DEP_WORBY_CANNATA_SUBP_000114696
NYC-WTC_000428379