Ambient Air Monitoring Program Section No. 10
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 56 of 116
extender and monitor body will be wrapped with tape to hold the cassette securely in place and
to provide a surface to identify the cassette. By inserting a pin through the hole near the display
screen, the elapsed time indicator is reset prior to sampling.
At the end of the sampling period, the top cover and small end caps are replaced. Samples are
transported upright to the laboratory in a rigid container with packing material to prevent jostling
or damage. To avoid electrostatic forces from causing fiber loss from the sampling filter,
untreated polystyrene foam will not be used for any of the shipping container materials.
Collection of a site-specific dust sample will be performed as discussed in Section 8.2.2. This
sample needs to be double-bagged prior to shipment to avoid any potential cross-contamination.
Equipment and supplies utilized with this sampling approach are as follows:
• Personal Monitor and Sampling Pump 0.5 to 3 1pm
• Battery Pack - Nickel metal hydride battery pack (or equivalent), fitted with a belt clip,
compatible with the personal sampling pump.
• Filters - Polyvinyl filter (or equivalent hydrophobic membrane filter) supported by a cassette
filter holder, 5.0-LLm pore size, suitable for connection with the personal monitoring pump
system.
• Cyclone - SKC Inc. aluminum cyclone (or equivalent)
• Medium Flow Rotometer - 0.4 to 4.0 1pm
• High Flow Rotometer - 2.0 to 20.0 1pm
• Forceps - Nylon, used when handling the filters
10.2.5 Polychlorinated Dibenzodioxins (PCDDs) and Polychlorinated Dibenzqfurans
(PCDFs)
Ambient air samples will be collected using a General Metals Works (GMW) PS-1 (or
performance equivalent) high volume air sampling system equipped for the collection of
semivolatile PCDD/PCDF particulate matter on a 102-mm diameter microquartz fiber filter, as
well as a glass cylinder containing a polyurethane sorbent (i.e., PUF plug) for the collection of
gas phase PCDDs/PCDFs. At each site, one GMW PS-1 sampler will be used to collect
PCDD/PCDF samples according to Method TO-9A. Samples will be collected at a flow rate of
200 to 300 liters per minute (1pm) over a sampling period of approximately 24 hours for a
resulting total air volume of approximately 288-432 cubic meters (288-432 m^).
L2005-346 Rev. 1
CITY DEP WORBY CANNATA SUBP 000114697
NYC-WTC_000428380