Part V - Quality Assurance Project Plan
133-135 Greenwich Street/21-23 Thames Street
Airtek Project 05-0701 - November 4, 2005
TABLE 13.4b
Summary of Calibration Procedures for Fixed Laboratory Analyses
Waste Sample Analysis - VOC, SVOC, Pesticides/Herbicides/PCB, Metals/ Mercury, pH
Instrument Activity Frequency Acceptance Corrective SOP
Criteria Action Ref.*
GCMSSVOC0117
DFTPP Tuning Every 12 hours SW846-8270C criteria Retune, rerun for 00
GC/MS Initial cal. (5 stds) As needed CCCs <30%RSD, SPCCs > DFTPP criteria
Semi Revision 1.4
method blank 1/20 per matrix 0.05 Rf; RSD avg.< 15% Recalibrate
Volatiles No targets found
Cont. calibration every 12 hours Re extract mblk
CCCs @ <25%D; SPCCs> Rerun or recahrerun all
Surrogate stds. ALL
0.05Rf afftected samples
Laboratory 1/20 per matrix
Within Lab control limits Re-extract
Control 1/20 per matrix
Within Lab control limits Reclibrate/ rerun all
MS/MSD ALL
samples
Internal Standards
-50 to +100% of 12 hour cal Reextract & rerun
chk std. Rerun affected samples,
dilute if matrix is issue
Hgl20998
Initial cal. (5 stds) As needed 0.995 Corr. Coeff. Recalibrate system
Merciiry/CV Revision 1.1
method blank Every 20 samples Not detected Rerun affected samples
AA
Cont. calibration Every 10 samples +/- 20% Rerun affected samples
Spike and Dup 1/20 samples Spike +/- 25%, Dup +/- 20% at Rerun spike/dup unless
levels >5x Reporting Limit proven matrix
Mfgs. Specs. interference
LCS 1/20 samples
Rerun all affected
samples
Recalibrate ICP system ICP031195
Initial calibration As needed +/- 10%
ICP metals Rerun affected samples Revision 1.2
Cont. calibration Every 10 samples +/- 10%
Rerun or qualify
Calibration blanks Every 10 samples Not detected (<IDL) affected samples
Calibrate with On use +- 0.05 units Recalibrate, if still > SOPWCpH
Corrosivity/ pH=7, and pH=4 0.05 replace probe 010596 Rev.1.6
buffers
pH