Part V - Quality Assurance Project Plan
133-135 Greenwich Street/21-23 Thames Street
Airtek Project 05-0701 - November 4, 2005
GCMSSVOCO11700
DFTPP Eveiy 12 SW846-8270C criteria Retune, rerun for
GC/MS Semi Tuning hours DFTPP criteria Revision 1.4
CCCs <30%RSD, SPCCs
Volatiles Initial cal. (5 As needed > 0.05 Rf; RSD avg.< 15% Recalibrate
stds) 1/20 per No targets found Re extract mblk
method blank matrix CCCs @ <25%D; SPCCs> Rerun or recal/rerun
Cont. every 12 0.05Rf all afftected samples
calibration hours Within Lab control limits Re-extract
Surrogate stds. ALL Within Lab control limits Reclibrate/ rerun all
Laboratoiy 1/20 per samples
Control matrix -50 to+100% of 12 hour Reextract & rerun
MS/MSD 1/20 per cal clik std. Rerun affected
Internal matrix samples, dilute if
Standards ALL matrix is issue
Hg 120998
Initial cal. (5 As needed 0.995 Corr. Coeff. Recalibrate system
Mercury/CVAA stds) Revision 1.1
Eveiy 20 Not detected Rerun affected
method blank samples +/- 20% samples
Cont. Eveiy 10 Spike +/- 25%, Dup +/- Rerun affected
calibration samples 20% at levels >5x samples
Spike and Dup 1/20 Reporting Limit Rerun spike/dup
samples Mfgs. Specs. unless proven matrix
interference
LCS
1/20 Rerun all affected
samples samples
Recalibrate ICP ICP031195
Initial As needed +/- 10%
system Revision 1.2
ICP metals calibration Every 10 +/- 10%
Rerun affected
Cont. samples Not detected (<IDL) samples
calibration Every 10 Rerun or qualify
Calibration samples
affected samples
blanks
Calibrate with On use +- 0.05 units Recalibrate, if still > SOPWCpH
Corrosivity/ pH=7, and 0.05 replace probe 010596 Rev.1.6
pH=4 buffers
pH
50
CITY_21MC100_003237454
NYC-WTC_000447483