A primary cell calibrator will be used to calibrate field sampling pumps. Calibration information
(i.e., date, times, equipment make/model/serial number, and results of calibration) will be
recorded in calibration logbooks.
Flow Rate Calibration - Prior to first use, or in the event that the sampling pump fails the flow
rate verification, the flow rate on the pump will be calibrated by setting the flow rate to the
respective set points using a suitable flow-measuring device that is NIST traceable.
PCDDs/PCDFs and PAHs using a PS-1 High Volume Sampler
The initial calibration of the PS-1 sampler will be performed using a critical orifice serving as a
reference standard in concert with a water manometer. The critical orifice calibration will be
performed by the manufacturer (Tisch) and is NIST traceable. The sampler will be calibrated
on-site using a certified flow orifice device before and after each sample collection period to
determine volumetric flow rates. All calibration measurements will be standardized to 760 mm
Hg and 25°C.
Table 13-3c summarizes the calibration procedures associated with the PS-1 samplers.
Table 13-3c. Calibration Requirements for PS-1 Samplers
Equipment Activity Acceptance Corrective Action
Criteria
PS-1 sampler Initial Calibration: NA NA
Multipoint calibration
performed before fust
use, quarterly thereafter,
after relocation to new
site, or after repairs
which may affect
calibration
Single point flow check Within ± 20% of the Service PS-1 sampler
at normal operating flow flow rate indicated by and perform a new
rate on monthly basis the PS-1 sampler multi-point
calibration.
Orifice Transfer Annual calibration by NA NA
Standard manufacturer and when
visual inspection reveals
new dents in the device.
PMi 0 andPMi s Using a Reference Sampler
Calibration procedures are as follows:
L2005-346 Rev. 1
C ITY_D E P_WO RB Y_C AN N ATA_SU B P_000114729
NYC-WTC_000428412