Ambient Air Monitoring Program Section No. 13
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 87 of 116
Table 13-3a. Calibration Requirements for High Volume Samplers
Equipment Activity Acceptance Corrective Action
Criteria
High volume sampler Initial Calibration: NA NA
Multipoint calibration
performed before first
use, quarterly thereafter,
after relocation to new
site, or after repairs
which may affect
calibration
Single point flow check Within ±20% of the Service high volume
at normal operating flow flow rate indicated by sampler and perform
rate on weekly basis the high volume a new multi-point
sampler calibration.
Orifice Transfer Annual calibration by NA NA
Standard manufacturer.
Mercury (Gaseous) Using an Ohio Lumex RA915+ Mercury Analyzer
Table 13-3b summarizes the calibration procedures associated with the mercury (gaseous)
analysis using the Ohio Lumex RA915+ Mercury Analyzer.
Table 13-3b. Calibration Requirements for the Ohio Lumex RA915+
Mercury Analyzer
Equipment Activity Acceptance Corrective Action
Criteria
Ohio Lumex RA Factory calibration prior NA NA
915+ to purchasing unit and
when unit continuously
fails operating
specifications
Calibration veritication: + 20% of true value Perform maintenance
performed prior to and and/or consult
after each sampling manufacturer
event (placed in Test
mode, internal test cell
containing mercury
vapor placed in optical
path of instrument)
L2005-346 Rev. 1