Ambient Air Monitoring Program Section No. 10
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 63 of 116
• Anderson RAAS PMio Sampler - A sampling system that meets the requirements of 40 CER
Part 50 Appendix J for the sampling of PMio when operated in the PMio mode. The
Anderson RAAS sampler design incorporates a flow-indicating device, a thermometer, a
barometer, and a timing control device into the sample monitor.
Manometer - A water or oil manometer is used to calibrate the flows of the PMio samplers.
• Timing/Control Device - A timing device capable of starting and stopping the sampler to
obtain an elapsed run time of 24 hr +1 hr. The accuracy of the timing device should be +15
minutes.
Flow Rate Transfer Standard - The flow rate transfer standard is a calibrated flow-metering
device traceable to NIST standards that is used to deliver a series of discrete flow rates into
the inlet of the PMio sampler so that the flow-indicating device on the sampler can be
accurately calibrated.
10.2.9.2 PMio - Continuous Monitoring ofPMw via Met One E-BAM Samplers
The monitors selected to continuously measure PMio are beta-attenuation monitors manufactured
by Met One Instruments, Inc. (Met One).
The E-BAM monitors will be equipped with particle size selective inlets. The design of the
inlets is such that particles larger than the desired size range will be removed from the air flow at
the prescribed air flow rate. The units will be equipped with an inlet head to separate PMio.
Sampling flow rate is critical to maintain the proper particle size cut points of the inlets. Plow
rates are maintained at 16.7 liters per minute in the E-BAM monitors using an integral flow
meter, pressure sensor, and ambient temperature sensor on board each monitor.
10.2.10 PM2.S
PM2.5 will be determined by two separate methods. At each location, PAL - will be determined
by a non-reference method utilizing Met One’s E-BAM Mass Monitor, which is a real-time
monitor. Anderson RAAS samplers will be used to collect PM2.5 via the US EPA standard
reference method (40 CER Part 50 Appendix L) at one location per day. An Anderson RAAS
PM2.5 sampler will be collocated along side the real-time PM2.5 monitors as a QA check. The
Anderson RAAS sampler will rotate on a monthly basis through all real-time PM2.5 monitor
locations for the duration of the monitoring program.
10.2.10.1 PM2.5 - Reference Method via Anderson RAAS Samplers
Particulate samples whose aerodynamic diameter is <2.5 p will be collected using Anderson
RAAS samples operating in the PAL - mode. All samples will be collected over an estimated
twenty-four hour sampling interval at a prescribed flow rate of 16.67 liters/minute. This equates