Ambient Air Monitoring Program Section No. 10
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 64 of 116
to a corrected air volume of or 24 m^ per each twenty-four hour sampling event. Preweighed
Teflon filters (47mm diameter) provided by the subcontracted laboratory will be placed in the
RAAS sampler cassette prior to the start of each sampling event.
The sampler specification for the PM2 5 sampler is contained in 40 CFR Part 50 Appendix L.
This document specifies the PM2.5 sampling limits, protocol, and instrument specifications. The
RAAS systems perform all of the functions required or recommended in the instrument
specification part of the standard. The inlet has 2.5-micrometer (nominal) cutpoint. The inlet
provides wind speed and direction independent sampling and removes particles larger than about
2.5 micrometers.
Before sampling, the filters are equilibrated to constant temperature and relative humidity
conditions and weighed. After sampling, the filters are again equilibrated to the constant
temperature and humidity conditions and weighed. The mass concentration is then calculated by
dividing the weight of the particulate captured on the filter by the volume of air (at ambient
conditions) that passed through the sampler. The flow rate is required to be maintained within
5% of 16.67 liters per minute with a coefficient of variation (a measure of the average short-term
variation in flow in %) of less than 2%.
Equipment and supplies utilized with this sampling approach are as follows:
Filters - Preweighed 47-mm Teflon® filters. Filters should be stored and transported in petri
dishes.
• Forceps - Non-serrated forceps should be used to handle the 47-mm filters.
• Anderson RAAS PM2.5 Sampler - A sampling system that meets the requirements of 40 CFR
Part 50 Appendix L for the sampling of PM2.5. The Anderson RAAS sampler design
incorporates a flow-indicating device, a thennometer, a barometer, and a timing control
device into the sample monitor.
• Manometer - A water or oil manometer is used to calibrate the flows of the samplers.
• Timing/Control Device - A timing device capable of starting and stopping the sampler to
obtain an elapsed run time of 24 hr ±1 hr. The accuracy of the timing device should be +15
minutes.
Flow Rate Transfer Standard - The flow rate transfer standard is a calibrated flow-metering
device traceable to NIST standards that is used to deliver a series of discrete flow rates into
the inlet of the PM2.5 sampler so that the flow-indicating device on the sampler can be
accurately calibrated.