The performance of the sample processing steps and the instrumentation are monitored on a
routine basis;' The procedures and ctiteria are summatized below.
One method blank and one laboratory spike sample are typically prepared with each ten samples of
any given matrix. The recoveries of the native PCDD and PCDF analytes in the laboratory spike
samples generally range from 70 to 130%. Recoveries of selected analytes outside this range do
not invalidate the data but provide information which is used by the laboratory to monitor recovery
trends and to assure optimization of the method.
Internal standards are spiked into each sample prior to extraction in order to monitor the level of
recovery which is achieved for each individual sample. Acceptable recoveries range from 40 to
135 percent unless a deviation is due to variation in instrument response as a result of analytical
interferences.
The resolution of the mass spectrometer is verified prior to each analysis to be 10,000 or greater.
Hardcopies of the reference peaks are printed at the beginning and end of each analysis day. The
resolving power of the DB-S chromatographic column is checked daily by analyzing a standard
;: solution containing 2,3,7,8-TCDD and the adjacent TCDD isomers. Acceptable performance is .
achieved when 2,3,7,8-TCDD is resolved from the adjacent isomers by a valley of 25% or less.
The group times for the selected-ion-monitoring data acquisitions are also checked daily by
analyzing the column performance mix which has been modified to contain the first and last eluting
isomers of each congener class. In this way one is assured of collecting data representative of the
total PCDD and PCDF content and that the 2,3,7,8-substituted isomers are suitably resolved.
Initial calibrations are generated by analyzing standard solutions (see Table 4) containing target
native and labeled PCDD and PCDF compounds. Response factors are calculated and averaged for
each compound. These averages are used for quantification and for comparison to the daily
continuing calibration. Acceptable ranges are specified in Method 23. A continuing calibration
standard is analyzed at the beginning of each 12-hour shift on days when initial calibrations are not
performed. The initial calibration is considered to be valid when the response factors from the
continuing calibration analysis agree to within the ranges specified in Method 23.