Quantification and Calculations (Cont.)
Each pair of ionjrnass peaks in the selected-ion-current chromatograms was evaluated manually to
determine if it'met the criteria for a PCDD or PCDF isomer. Areas, of all peaks exhibiting correct ion
ratios and havTng retention times within the correct windows were then summed for calculations of
total congener concentrations. The 2,3,7,8-TCDD toxic equivalency factors (TEFs) used to
calculate the sample toxic equivalency are listed in Table 5,
, 2'"'"' ■■ ■ ■ '
A iimit of detection (LOD) based on producing a signal that is 2.5 times the noise level, was
calculated for each undetected 2,3,7,8-substituted isomer. The noise heights used to calculate the
detection limits were measured at the retention time of the specific isomer. The formula used for
calculating the LOD is: .
Hn X Qis X 2.5
LOD =
His X Rf
where:
LOD = Single isomer limit of detection
Hn = Sum of noise heights at native isomer retention time
Qis = Quantity of labeled internal standard
His = Sum of peak heights for labeled internal standard
Rf. = Response factor
The recoveries of the labeled surrogate compounds, relative to the corresponding internal standard,
and the ^^Cij-labeled internal standards, relative to either 1,2,3,4-TCDD-''2Ci2 or 1,2,3,7,8,9-
j; HxCDD-^Ct2, were calculated using the follovving equation: . \ . ?
Ais X Ors X 100%
%R =
Rfr X Ars x Qis
where:
%R = Percent recovery of labeled internal standard
Ais Sum of integrated areas of labeled internal standard
\\Qrs Quantity of recovery standard
Ars Sum of integrated areas of recovery standard
Rfr Response factor of the specific labeled internal standard relative to the recovery standard
Qis Quantity of the labeled internal standard added to the sample
REPORT OF LABORATORY ANALYSIS