Quantification and Calculations
The PCDD and P.GOF isomers were quantified by comparison of their responses to the responses of
the labeled inter;nal standards as described in USEPA Method 23.. Relative response factors were
calculated . from analyses of standard mixtures containing representatives of each of the
PCDD/PCDF congener classes at five concentration levels, and each of the internal standards at
one concentration level, as shown in Table 4. The PCDD/PCDF response factors were calculated
by/comparing the sum of the responses from the two ion masses monitored for each chlorine
congener class to the sum of the responses from the two ion masses of the corresponding
isotopically labeled internal standard. The formula for the response factor calculation is:
An X Qis
Rf =
Ais X Qn
where:
Rf Response factor
An Sum of integrated areas for native isomer
Qis Quantity of labeled internal standard
Ais Sum of integrated areas for labeled internal standard
Qn Quantity of native isomer :
The levels of PCDDs and PCDFs in each sample were quantified using the following equation:
An X Qis
r C =
Ais X Rf
where:
C Concentration of target isomer or congener class
An Sum of integrated areas for the target isomer or congener class
Qis Quantity of labeled internal standard added to the sample
Ais Sum of integrated areas for the labeled internal standard ,
■■ . Rf Response factor
REPORT OF LABORATORY ANALYSIS
- . This report shall not be reproduced,’except in full,
without the written consent of Pace Analytical Services, Inc.