NYC-WTC_000053094
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 8, 2002.
1 shared topics · Same agency, volume and box · similarity 0.999
machine-extracted · confidence unavailable · Check page ↗Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 14, 2002.
NYC-WTC_000053389–000053503Folder label: “ICPMS 5/14/02”

OCR status: ok · source: pdftotext
NYC 9/11 Public Portal Document
WESTCHESTER COUNTY DEPT OF LABS AND RESEARCH Method 200.8 - Summary Report Sample ID: AE09577 Sample Date/Time: Tuesday, May 14, 2002 09:53:20 Sample Type: Spike -1 Sample Description: Number of Replicates: 3 Method File: c:\winnt\profiles\lrd1\elandata\Method\NEWQ200.8.mth Dataset File: c:\winnt\profiles\lrd1\elandata\Dataset\newyork city\AE09577.012 Aliquot Volume (mL): Diluted To Volume (mL):
Concentration Results Analyte Mass Meas. Intens. Mean Meas. Intens. RSD Net Intens. Mean Cone. Mean Cone. RSD Sample Unit Be 9 24680.706 1.068 0.042' 51.55903 2.92 ug/L Se-1 45 582736.743 2.122 582736.743 ug/L Cr 52 623839.609 3.004 1.055 49.37000 0.97 ug/L Cr 53 76617.948 0.528 0.130 51.25989 1.66 ug/L Mn 55 1478503.716 1.973 2.535 70.64056 0.57 ug/L Ni 60 158355.447 1.664 0.272 51.28605 0.93 ug/L Cu 63 382835.221 1.217 0.657 58.08384 1.80 ug/L Cu 65 185158.032 1.521 0.317 58.68552 0.79 ug/L Zn 66 120547.956 1.427 0.925 54.50344 2.13 ug/L Zn 67 21038.198 1.963 0.161 54.12412 0.97 ug/L Zn 68 91319.310 1.100 0.700 55.34910 0.66 ug/L Ge 72 129256.099 1.123 129256.099 ug/L As 75 144437.005 0.252 1.073 52.32703 1.36 ug/L Se 77 14841.709 0.329 0.108 56.90626 1.14 ug/L Se 82 16228.648 1.102 0.126 55.52556 0.95 ug/L Kr 83 2318.319 2.538 -186.385 ug/L Ag 107 664145.195 0.898 0.393 45.70212 1.55 ug/L > In 115 1690907.453 0.922 1690907.453 ug/L Sb 121 572702.165 0.663 0.339 54.80023 0.79 ug/L Sb 123 441022.997 1.436 0.261 55.50719 1.62 ug/L Ba 135 220999.346 1.029 0.083 64.17463 1.56 ug/L Ba 137 387273.279 0.709 0.145 63.94066 0.90 ug/L > Tb 159 2677351.076 1.412 2677351.076 ug/L Hg 201 6735.562 0.631 0.002 4.01413 0.84 ug/L Hg 202 15387.500 0.967 0.006 4.01742 0.79 ug/L U 205 1757415.820 2.077 0.656 43.62227 0.95 ug/L Pb 208 2367406.798 0.407 0.883 47.34036 1.20 ug/L
QC Calculated Values Analyte Mass QC Std % Recovery Int Std % Recovery Spike % Recovery Dilution % Diff Dup. Rel. % Diff ' Be 9 103.118 > Sc-1 45 82.693 Cr 52 98.549 Cr 53 101.168 Mn 55 100.690
Sample ID: AE09577 Report Date/Time: Tuesday, May 14, 2002 09:55:42 Page 1
NYC-WTC_000053412
OCR can misread numbers and units. Confirm readings against the page image before using them.
Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.
NYC-WTC_000053094
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 8, 2002.
1 shared topics · Same agency, volume and box · similarity 0.999
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053337
Daily instrument calibration and performance verification log for ICPMS equipment conducted on April 25, 2002.
1 shared topics · Same agency, volume and box · similarity 0.998
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040925
Summary report of water sample analysis using Method 200.8 showing measured intensities and concentrations for various metals.
1 shared topics · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040700
Summary report of water sample analysis using Method 200.8 showing measured intensities and concentrations for various metals.
1 shared topics · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040648
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040871
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040755
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040341
Summary report of water sample analysis using Method 200.8 showing measured intensities and concentrations for various metals.
1 shared topics · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040452
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.995
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040598
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.995
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040287
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.995
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000078316
Laboratory report from Westchester County Dept of Labs and Research detailing Method 200.8 metal concentrations in water samples dated April 8, 2002.
1 shared topics · similarity 0.993
machine-extracted · confidence unavailable · Check page ↗Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.
Page similarity is temporarily unavailable or this page has no indexed vector.
Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.
NYC-WTC_000053388
Administrative record from DEP Box 29 documenting ICPMS activities at the World Trade Center site in May 2002.
Filed in the same folder
NYC-WTC_000053094
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 8, 2002.
Similar indexed content
NYC-WTC_000053337
Daily instrument calibration and performance verification log for ICPMS equipment conducted on April 25, 2002.
Similar indexed content
NYC-WTC_000040925
Summary report of water sample analysis using Method 200.8 showing measured intensities and concentrations for various metals.
Similar indexed content
NYC-WTC_000040700
Summary report of water sample analysis using Method 200.8 showing measured intensities and concentrations for various metals.
Similar indexed content