Daily Performance Report
Sample ID: DPC
Sample Date/ Time: Tuesday, April 09, 2002 09:49:22
Sample Description: DAILY PERFORMANCE CHECK
Method File: d:\elandata\Method\DAILY PERF TO SAVE REPORT.mth
Dataset File: d:\elandata\Dataset\Daily\DPC.040
Tuning File: d:\elandata\Tuning\default.tun
Optimization File: d:\elandata\Optlmize\default.dac
Dual Detector Mode: Pulse
Acq. Dead Ti► e(ns): 55
Current Dead Time (ns): 55
Summary
Analyte Mass Meas. Intens. Mean Net Intens. Mean Net Intens. SD Net Intens. RSD
Mg 24.0 52005.1 52005.0761 634.912 1.2
Rh 102.9 306331.9 306331.857 ✓ 3396.367 1.1
In 114.9 426359.4 426359.419 l 6748.404 1.6
Pb 208.0 233036.6 233036.587 A 2596.011 1.1
1> Ba 137.9 385183.4 385183.387 / 5369.931 1.4
L Ba++ 69.0 5109.5 0.013 d 0.000 0.9
(> Ce 139.9 473473.9 473473.885 / 5690.803 1.2
L CeO 155.9 13105.4 / 0.028 0.000 1.2
Bkgd 220.0 11.0 J 11.000 1.196 10.9
Current Optimization File Data
Current Value Description
0.88 Nebulizer Gas Flow
6.00 Lens Voltage
1000.00 ICP RF Power
-1900.00 Analog Stage Voltage
1600.00 Pulse Stage Voltage
70.00 Discriminator Threshold
-6.00 AC Rod Offset
60.00 Service DAC I
0.00 Quadrupole Rod Offset
Current Autolens Data
Analyte Mass Num of Pis DAC Value Maximum Intensity
Be 9 21 5.3 5796.8
Co 59 21 6.0 161848.0
In 115 21 6.5 475180.2