NYC-WTC_000053094
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 8, 2002.
1 shared topics · Same agency, volume and box · similarity 0.998
machine-extracted · confidence unavailable · Check page ↗Daily instrument calibration and performance verification log for ICPMS equipment conducted on April 25, 2002.
NYC-WTC_000053337–000053387Folder label: “ICPMS 4/25/02”

OCR status: ok · source: pdftotext
NYC 9/11 Public Portal Document
WESTCHESTER COUNTY DEPT OF LABS AND RESEARCH Method 200.8 - Summary Report Sample ID: AE08960 Sample Date/Time: Thursday, April 25, 2002 14:54:44 Sample Type: Spike -1 Sample Description: Number of Replicates: 3 Method File: c:\winnt\profiles\lrd1\elandata\Method\NEWQ200.8.mth Dataset File: c:\winnt\profiles\lrd1\elandata\dataset\newyork city\AE08960.700 Aliquot Volume (mL): Diluted To Volume (mL):
Concentration Results Analyte Mass Meas. Intens. Mean Meas. Intens. RSD Net Intens. Mean Cone. Mean Cone. RSD Sample Unit Be 9 18397.114 0.903 0.040 53.96238 0.59 ug/L > Se-1 45 454067.653 1.425 454067.653 ug/L Cr 52 412792.923 2.020 0.894 48.45773 1.55 ug/L Cr 53 48763.592 0.773 0.106 48.49723 1.41 ug/L Mn 55 839871.712 2.192 1.846 64.13418 2.45 ug/L Ni 60 99303.264 2.333 0.218 47.72588 3.12 ug/L Cu 63 267001.780 1.059 0.587 56.57561 2.36 ug/L Cu 65 131217.219 0.878 0.289 56.16898 1.58 ug/L Zn 66 79443.815 1.847 0.878 52.23964 2.07 ug/L Zn 67 14308.061 1.092 0.158 51.48920 2.25 ug/L Zn 68 60061.021 0.703 0.664 52.65077 1.71 ug/L > Ge 72 89391.608 1.152 89391.608 ug/L As 75 91457.667 0.683 1.022 50.78819 1.65 ug/L Se 77 9059.693 2.358 0.100 55.61253 1.49 ug/L Se 82 10689.389 2.010 0.118 53.35043 1.32 ug/L Ag 107 441811.649 0.491 0.372 45.95165 0.61 ug/L > In 115 1185915.413 0.147 1185915.413 ug/L Sb 121 411457.693 0.234 0.347 51.47657 0.38 ug/L Sb 123 318448.176 1.175 0.269 51.32708 1.27 ug/L Ba 135 160213.524 ' 1.374 0.099 65.57110 1.68 ug/L Ba 137 279093.506 0.494 0.172 65.41972 1.06 ug/L Ho 165 1601757.891 1.499 -0.002 ug/L Tb 159 1623440.252 0.599 1623440.252 ug/L Hg 201 4191.342 2.914 0.003 4.08622 2.42 ug/L Hg 202 9770.345 0.544 0.006 4.19864 1.00 ug/L TI 205 1053763.288 0.349 0.649 48.94768 0.79 ug/L Pb 208 1426748.869 1.225 0.878 49.17236 1.52 ug/L
QC Calculated Values Analyte Mass QC Std % Recovery I nt Std % Recovery Spike % Recovery Dilution % Diff Dup. Rei. % Diff ’ Be 9 107.935 > Sc-1 45 108.280 Cr 52 96.599 Cr 53 96.308 Mn 55 95.907
Sample ID: AE08960 Report Date/Time: Thursday, April 25, 2002 14:56:58 Page 1
NYC-WTC_000053360
OCR can misread numbers and units. Confirm readings against the page image before using them.
Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.
NYC-WTC_000053094
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 8, 2002.
1 shared topics · Same agency, volume and box · similarity 0.998
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000053389
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 14, 2002.
1 shared topics · Same agency, volume and box · similarity 0.998
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040925
Summary report of water sample analysis using Method 200.8 showing measured intensities and concentrations for various metals.
1 shared topics · similarity 0.998
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040871
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040598
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040648
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040700
Summary report of water sample analysis using Method 200.8 showing measured intensities and concentrations for various metals.
1 shared topics · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040755
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.997
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040341
Summary report of water sample analysis using Method 200.8 showing measured intensities and concentrations for various metals.
1 shared topics · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040452
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000040287
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
1 shared topics · similarity 0.996
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000077621
Laboratory report from Westchester County Dept of Labs and Research detailing Method 200.8 metal concentrations in water samples dated March 25, 2002.
1 shared topics · similarity 0.994
machine-extracted · confidence unavailable · Check page ↗Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.
Page similarity is temporarily unavailable or this page has no indexed vector.
Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.
NYC-WTC_000053336
Administrative record from DEP Box 29 documenting ICPMS activities at the World Trade Center site in April 2002.
Filed in the same folder
NYC-WTC_000053094
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 8, 2002.
Similar indexed content
NYC-WTC_000053389
Daily instrument calibration and performance verification log for ICPMS equipment conducted on May 14, 2002.
Similar indexed content
NYC-WTC_000040925
Summary report of water sample analysis using Method 200.8 showing measured intensities and concentrations for various metals.
Similar indexed content
NYC-WTC_000040871
Daily instrument performance verification log for ICPMS analysis including intensity and standard deviation data for multiple elements.
Similar indexed content