GCMS calibration data for sample AE07484, May 2002

Machine-extracted title · confidence 90%

Calibration report listing quantified values for various organic compounds including nitrosamines and chlorobenzenes used for instrument verification.

NYC-WTC_000044253–000044256

Folder label: “GCMS X 4/23/02 A 4/26/02

Page image
Scanned page image, NYC-WTC_000044255
OCR text

OCR status: ok · source: pdftotext

NYC 9/11 Public Portal Document

Quantitation Report (QT Reviewed)

Data File C:\HPCHEM\1\DATA\APR2 5 02\CAL50A.D Vial: 2 Acq On 26 Apr 2002 10:53 am Operator : Sample Inst : 209 Mise Multiplr: 1.00 MS Integration Params: GOOFY.P Quant Time: Apr 26 12:38 2002 Quant Results File: GCMS0412.RES

Quant Method : C:\HPCHEM\1\METHODS\GCMS0412.M (RTE Integrator) Title : 209 625/TCLP calibration table Last Update : Thu Apr 25 15:21:43 2002 Response via : Initial Calibration DataAcq Meth : GCMS0412

Compound R.T. Qlon Response Cone Unit Qvalue

33) Hexachlorobenzene 24.70 284 761476 54.56 ug/1 95 34) Phenanthrene 25.67 178 3125824 49.81 ug/1 100 35) Anthracene 25.81 178 3096973 50.17 ug/1 100 36) Di-n-butylphthalate 27.57 149 3809247 48.52 ug/1 100 37) Fluoranthene 29.30 202 3182283 50.41 ug/1 # 92 4 0) Pyrene 29.97 202 3199694 47.86 ug/1 97 41) Butylbenzylphthalate 32.09 149 1464786 44.80 ug/1 99 42) Benzo(a)anthracene 33.60 228 2340963 50.48 ug/1 100 43) Bis(2-ethylhexyl)phthalate 33.87 149 1796495 42.94 ug/1 # 99 44) Chrysene 33.74 228 2261520 49.17 ug/1 99 46) Di-n-Octylphthalate 35.62 149 2692403 43.17 ug/1 # 98 47) Benzo(b)fluoranthene 36.72 252 1935230 48.03 ug/1 98 48) Benzo(k)fluoranthene 36.79 252 2065022 52.63 ug/1 97 4 9) Benzo(a)pyrene 37.70 252 1673585 50.38 ug/1 99 50) Indeno(1,2,3-cd)pyrene 42.00 276 1591066m 47.57 ug/1 51) Dibenz(a,h)anthracene 42.06 278 1233379 48.36 ug/1 96 52) Benzo(g,h,i)perylene 43.24 276 1246909 44.66 ug/1 96

(#) = qualifier out of range (m) = manual integration CAL50A.D GCMS0501.M Wed May 01 11:18:25 2002 Page 2

NYC-WTC_000044255

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000044255Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Page similarity is temporarily unavailable or this page has no indexed vector.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

Compare copies and versions →