PROCEDURES OPERATION AND MAINTENANCE OF THE LANDFILL GAS PRODUCTION FACILITY AT FRESH KILLS LANDFILL

FKP-0019: LFG WELLFIELD - TEST EQUIPMENT CALIBRATION

Purpose

Wellfield optimization requires accurate LFG quality data. To ensure the data collected in the field is accurate, proper instrument calibration is required.

Equipment

• Certified Methane, Oxygen, and Carbon Dioxide calibration gases

Procedure

For instrument specific calibration instructions, refer to manufacturers recommended procedures. For all calibration procedures, calibration must employ a minimum of two calibration gases, a zero and a span (high end). The span gas should be within 20 to 50 percent of the regulatory limit (if one applies). Therefore, the span gas for oxygen calibration should be 4% oxygen. When calibrating for methane, ambient air can be used as a zero gas provided calibration is performed upwind from any methane sources. Some suggested calibration gases are as follows.

• For oxygen; zero can be set using a 15% CO2/15% CH4 calibration gas (balance nitrogen), and the span can be accomplished with a 4% O2 (balance nitrogen).

• For methane, zero can be set with ambient air, span can be set with either the 35% 002/50% CH4 calibration gas (balance nitrogen) or a 15% 002/15% CH4 calibration gas (balance nitrogen). The expected readings will dictate the appropriate calibration gas (i.e., if readings at the wellheads are anticipated to be approximately 55% GH4, the 50% GH4 span gas would be used).

• For carbon dioxide, the span can be set using 35% 002/50% CH4 calibration gas (balance nitrogen), and the zero can be accomplished with a 4% O2 (balance nitrogen).

• For methane surface emissions monitoring, the FID shall be calibrated with a 500 ppm methane calibration gas.

1

CITY_21MC100_010050410

NYC-WTC_000454384

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000454384Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

NYC-WTC_000421711

Quality assurance plan for ambient air monitoring during LMDC project at 130 Liberty Street.

Similar indexed content

NYC-WTC_000428324

Draft QA project plan for Ambient Air Monitoring Program at 130 Liberty Street deconstruction (Aug 2005).

Similar indexed content

NYC-WTC_000427221

Draft QA Project Plan for ambient air monitoring at 130 Liberty Street deconstruction, TRC scope, Aug 2005.

Similar indexed content

NYC-WTC_000449648

Draft QA Plan for ambient air monitoring at 130 Liberty Street deconstruction project, Aug 2005.

Similar indexed content

Compare copies and versions →