detailed in the associated methods and in Tables 8-2 through 8-10.
11.2.2.1 Method Blanks/Preparation Blanks Method blanks will be performed as part of
each analytical batch for each methodology performed. Method blanks are used to
evaluate contamination introduced during sample preparation and/or analysis by the
laboratory.
11.2.2.2 Instrument Blanks Instrument blanks are used to evaluate contamination
resulting from the analytical reagents and the instrumentation. In addition, instrument
blanks are sometimes used to assess potential carryover after the analysis of a highly
contaminated sample. Instrument blanks are only required for select analytical
parameters.
11.2.2.3 Surrogate Spikes Surrogate spikes are used to evaluate extraction efficiency or
analytical bias on a sample by sample basis for organic parameters. Surrogate spikes are
added to all samples for organic parameters. Surrogate spikes are another measure of
sample-specific QC.
11.2.2.4 Laboratory Control Samples Laboratory control samples (LCSs) and LCS
Duplicates are used to evaluate almost all parameters for the ability of the laboratory to
accurately and precisely identify and quantitate target compounds in a reference matrix
when spiked at the mid range of the calibration curve at a known concentration using a
secondary source standard. LCSs and/or LCS Duplicates are typically performed as part
of each analytical batch for each methodology with the exception of asbestos, PMio,
PM2.5, ignitability and corrosivity.
11.2.2.5 Laboratory Duplicate Laboratory duplicates are used to evaluate laboratory
preparation and analysis precision. These analyses are typically performed for inorganic
parameters only. Laboratory duplicates are typically performed at a frequency of one per
twenty samples.
1 L2.2.6 Internal Standards Internal standards are used to assess the analytical accuracy,
precision, and stability. Internal standards are typically only used for organic analyses
and ICP/MS analyses. Internal standards are spiked into all samples and are considered a
sample-specific QC measure.
11.2.2.7 Standard Reference Materials Standard reference materials (SRMs) are used to
evaluate laboratory preparation and analysis bias for specific compounds in a reference
matrix. SRMs will be used for the asbestos analysis.
11.2.2.8 Matrix Spike Samples The matrix spike samples are used to determine laboratory
preparation and analysis bias for specific compounds in specific matrices (i.e., sample
specific QC). Matrix spikes are typically performed at a frequency of one per twenty
investigative samples per analytical parameters with the exception of the RCRA
characteristics.