Part V - Quality Assurance Project Plan
133-135 Greenwich Streel/21-23 Thames Street
Airtek Project 05-0701 - Rev 1 - February 1, 2006
Grid Opening Variation of grid
Measurement: openings must not
performed on 2 % of be > 5% of the mean
grid lot to determine
average grid
opening in mm^
Silica XRD Standard: daily 80-120% of true Perform necessary
analysis value equipment
maintenance and
Initial Calibration: Acceptable daily
check calibration
whenever there is a standard analysis
standards.
change in hardware results calculated
or when a new set of using this curve;
calibration reference intercept value less
standards are than 5 pg for
prepared primary peak curves
PM 10 Balance Calibration: ±3 pg Perform necessary
daily before use and equipment
every 10 samples maintenance and
(200,000 mg) check calibration
Calibration ±3 pg standards.
Verification: daily
before use (100, 000
mg)
Tab/e 13-4a. Summary of Ca/ibration Procedures for Fixed Laboratory Ana/j/ses
Parameter Frequency of Acceptance Corrective Action
Instrument Calibration Criteria
PM 25 Balance Calibration: ±3 pg Perform necessary
daily before use and equipment
every 10 samples maintenance and
(200,000 mg) check calibration
Calibration ±3 pg standards.
Verification: daily
before use (100, 000
mg)
TABLE 13.4b
Summary of Calibration Procedures for Fixed Laboratory Analyses
Waste Sample Analysis - VOC, SVOC, Pesticides/Herbicides/PCB, Metals/ Mercury, pH
Instrument Activity Frequency Acceptance Corrective SOP
Criteria Action Ref.*
102