Part V - Quality Assurance Project Plan
133-135 Greenwich Streel/21-23 Thames Street
Airtek Project 05-0701 - Rev 1 - February 1, 2006
Retune GCMSVOC01170
BFB Tuning Every 12 hours Method 8260B Criteria
Recalibrate system 0
GC/MS Initial cal. (5 stds) As needed CCC RSDs < 30%’ Rf for
Volatiles Every 12 hours Rerun blank, rerun Revision 1.7
Method blank SPCCs >0.05, RSD Avg <
Every 12 hours 15% affected samples
Cont. calibration
All Acetone, McCL < 25 ppb Rerun cal chk and all
Surrogate stds.
CCCs < 25%D; SPCCs > 0.05 affected samples
Laboratory Oncc/20 samples
Within Laboratory Control Rerun affteced samples
Control Once/20 samples
limits Rerun LCS and all
MS/MSD ALL
affected samples
Internal Standards
-50% to +100% of 12 hr Cal. Rerun MS/MSD and
affected samples
Rerun affected samples
Recalibrate System GCPEST011799
Initial cal. (5 stds) As needed RSDs all <20%
GC Cont. calibration Rerun all affected Revision 1.1
Every 10 samples < 15 % Difference
Pesticides/ (midpoint) samples under good GCHRRBOl 1999
Herbicides method blank
one per batch Not detected cev Revision 1.2
and PCBs All Within Lab control limits Reextract, rerun
surrogate stds. GCPCB011799
Once per 20 Within lab control limits Re extract samples with
MS/MSD & LCS Revision 1.2
samples both surrogates out
Reextract or reanalyze
affected samples where
LCS is out.
GCMSSVOC0117
DFTPP Tuning Every 12 hours SW846-8270C criteria Retune, rerun for
00
GC/MS Semi- Initial cal. (5 stds) As needed CCCs <30%RSD, SPCCs > DFTPP criteria
Volatiles Revision 1.4
method blank 1/20 per matrix 0.05 Rf; RSD avg.< 15% Recalibrate
No targets found Re extract mblk
Cont. calibration every 12 hours
CCCs @ <25%D; SPCCs> Rerun or recal/reiun all
Surrogate stds. ALL
0.05Rf afftected samples
Laboratory 1/20 per matrix
Within Lab control limits Re-extract
Control 1/20 per matrix
Within Lab control limits Reclibrate/ rerun all
MS/MSD ALL
samples
Internal Standards
-50 to +100% of 12 hour cal Reextract & rerun
chk std. Rerun affected samples,
dilute if matrix is issue
Hgl20998
Initial cal. (5 stds) As needed 0.995 Corr. Coeff. Recalibrate system
Mercuiy/CV Revision 1.1
method blank Every 20 samples Not detected Rerun affected samples
AA
Cont. calibration Every 10 samples +/- 20% Rerun affected samples
Spike and Dup 1/20 samples Spike +/- 25%, Dup +/- 20% at Rerun spike/dup unless
levels >5x Reporting Limit proven matrix
Mfgs. Specs. interference
LCS 1/20 samples
Rerun all affected
samples
Recalibrate ICP system 1CP031195
Initial calibration As needed +/- 10%
ICP metals Rerun affected samples Revision 1.2
Cont. calibration Every 10 samples +/- 10%
Rerun or qualify
Calibration blanks Every 10 samples Not detected (<1DL)
affected samples