Table 13-lb, Maintenance, Testing, and Inspection Requirements for the Ohio
Lumex RA915+ Mercury Analyzer
Fault Symptom Possible Cause Corrective Action
The battery discharge indicator The battery is fully discharged Charge the battery
(red) glows for some time and
then goes out when the analyzer
is switched on.
In the TEST mode, relative - Spectral lamp is not switched - Press the "T.amp ignition"
deviation R of the measured test on button
number differs from the
tabulated value by more than - The test cell switch is in idle - Set the switch of the test cell
25 %. position to the working position
- The test cell is out of order - Shake the test cell 2-3 times
by the TEST (ON/OET) switch
- The absorption filter has failed - Replace the filter
Asbestos, Mercury, and PCBs using a Personal Sampling Pump
Maintenance, testing and inspection requirements for the personal sampling pumps are as
follows:
• Routine maintenance will be performed according to the manufacturer’s instructions on the
following schedule:
_______ Part Replacement Interval (hours)
Valves 2,000
Pump Diaphragm 2,500
Dampner 2,500
Motor 4,500
Note: Replacement intervals may be shorter in dusty conditions.
• Flow Rate Verification - On a weekly basis, the field technician will verify the sampling
flow rate on the sampling pump using a rotameter, bubble meter, or other suitable flow
measuring device that is NIST traceable. The flow rate should not vary from the set point by
more than ±10%. If it does, the flow through the pump must be recalibrated.
Respirable Silica using a Personal Sampling Pump
Maintenance, testing, and inspection requirements for the personal sampling pump are as
follows:
L2005-346 Rev. 1
C ITY_D E P_WO RB Y_C AN N ATA_SU B P_000114723
NYC-WTC_000428406