Ambient Air Monitoring Program Section No. 13
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 81 of 116
Table 13-la. Maintenance, Testing, and Inspection Activities Associated
with Metals (TSP filters) Collected with High Volume Samplers
Acceptance
Equipment Activity Corrective Action
Criteria
Brushes Replace after 600 - Stable flow rate Replace as necessary
1000 hrs. of operation
Motor Replace as needed Consult manufacturer Obtain the correct
for correct model of model
motor
Tubing and fittings Visually inspect daily No crimps, cracks, or Replace as necessary
obstructions; no
crosstlireading
Mercury (Gaseous) Using a Direct Read Analyzer
Table 13-lb summarizes the inspection, testing, and maintenance activities associated with the
gaseous mercury analysis using the Ohio Lumex RA915-I- mercury analyzer.
Table 13-lb. Maintenance, Testing, and Inspection Requirements for the Ohio
Lumex RA915+ Mercury Analyzer
Fault Symptom Possible Cause Corrective Action
Segments of the indicator table - Power cable is out of -Repair the power cable
on the display and control unit order
are not highlighted when the
analyzer is switched on. - Display unit cable is - Repair the display unit
out of order cable
- Battery is discharged - Charge the battery
The (*) symbol at the display - Battery is discharged - Charge the battery
and control panel is not dimmed
out when the "Lamp ignition"
button is pressed:
-if the optical switch is in - Operation is possible only with -Set the optical switch into
position I attachments positions 11 or 111
- if the optical switch is in - The single-pass cell is - Remove the single-path cell,
position 11 contaminated, compartment check if the compartment
windows are contaminated or windows are clean, make sure
foreign objects are found in the that there are no foreign objects
compartment inside tlie compartment
- if the optical switch is in - Multi-path cell is - Clean the multi-path cell
position 111 contaminated (see appropriate section)