13.0 TESTING, INSPECTION, MAINTENANCE AND CALIBRATION
REQUIREMENTS
13.1 Instrument/Equipment Testing, Inspection, and Maintenance
13.1.1 Field Equipment
This section describes the procedures and documentation activities that will be performed to
ensure that all field analytical instrumentation and equipment are available and in working order
when needed. Instrument maintenance logs must be kept and instrumentation must be checked
prior to use. The field instrument preventative maintenance program is designed to ensure the
effective completion of the sampling effort and to minimize instrument downtime. The
maintenance responsibilities for field instruments will be assigned to the TRC Field Sampling
Coordinator. Field personnel will be responsible for daily field checks and calibrations and for
reporting any problems with the instruments. The maintenance schedule will follow the
manufacturer’s recommendations. Field personnel will also be responsible for ensuring that
critical parts are included with the field instruments. Critical spare parts will be immediately
available to reduce potential downtime. The inventory will primarily contain parts that are
subject to frequent failure, have limited useful lifetimes, and/or cannot be obtained in a timely
manner.
A spare set of equipment will be maintained on-site to facilitate network continuous operation in
the event of equipment failure that cannot be remedied by in situ repair. Additional instruments
and equipment will be available within 1-day shipment to avoid delays in the field schedule.
Metals (TSP Filters)
Table 13-la summarizes the inspection, testing, and maintenance activities associated with the
high volume samplers used to collect metals (TSP) samples.
Table 13-la, Maintenance, Testing, and Inspection Activities Associated
with Metals (TSP filters) Collected with High Volume Samplers
Acceptance
Equipment Activity Corrective Action
Criteria
High volume sampler Single point flow check Within ±20% of the Service high volume
at normal operating flow flow rate indicated by sampler and perform
rate weekly- the high volume a new multi-point
sampler calibration
Power cords Check for crimps or No obvious damage Replace as necessary
cracks
Filter screen and Visually check daily No obvious deposits Wipe surface clean
throat during sample recovery as needed
procedure
Gaskets At 3 month intervals, No leaks or Replace as necessary
inspect all gaskets in tlie compression damage
sampler evident