Ambient Air Monitoring Program Section No. 11
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 70 of 116
11.2.2 Fixed Laboratory QC
All required QC checks and QC samples and the associated QC acceptance limits are detailed in
the associated methods and in Tables 8-2 through 8-9.
11.2.2.1 Method Blanks/Preparation Blanks
Method blanks will be performed as part of each analytical batch for each methodology
performed. Method blanks are used to evaluate contamination introduced during sample
preparation and/or analysis by the laboratory.
11.2.2.2 Instrument Blanks
Instillment blanks are used to evaluate contamination resulting from the analytical reagents and
the instrumentation. In addition, instrument blanks are sometimes used to assess potential
carryover after the analysis of a highly contaminated sample. Instrument blanks are only
required for select analytical parameters.
11.2.2.3 Surrogate Spikes
Surrogate spikes are used to evaluate extraction efficiency or analytical bias on a sample by
sample basis for organic parameters. Surrogate spikes are added to all samples for organic
parameters. Surrogate spikes are another measure of sample-specific QC.
11.2.2.4 Laboratory Control Samples'
Laboratory control samples (LCSs) and LCS Duplicates are used to evaluate almost all
parameters for the ability of the laboratory to accurately and precisely identify and quantitate
target compounds in a reference matrix when spiked at the mid range of the calibration curve at a
known concentration using a secondary source standard. LCSs and/or LCS Duplicates are
typically performed as part of each analytical batch for each methodology with the exception of
asbestos, PMio and PM2.5.
11.2.2.5 Laboratory Duplicate
Laboratory duplicates are used to evaluate laboratory preparation and analysis precision. These
analyses are typically performed for inorganic parameters only. Laboratory duplicates are
typically performed at a frequency of one per twenty samples.
11.2.2.6 Internal Standards
Internal standards are used to assess the analytical accuracy, precision, and stability. Internal
standards are typically only used for organic analyses and ICP/MS analyses. Internal standards
are spiked into all samples and are considered a sample-specific QC measure.