Ambient Air Monitoring Program Section No. 10
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 60 of 116
10.2.7 Polycyclic Aromatic Hydrocarbons (PAHs)
PAH samples will be collected using high-volume air samplers fitted with non-size selective
quartz fiber filters and sorbent cartridges. This approach, as described in EPA Method TO-13A,
does not collect particulate matter representative solely of potential inhalation exposure, but
provides for a total of respirable and non-respirable PAHs. Sorbent traps containing
polyurethane foam (PUP) and XAD-2® resin in a “sandwich” configuration. The General Metals
Works (GMW) PS-1 high volume sorbent sampler (or performance equivalent) located at each
monitoring station will be operated at a flow rate of 200 to 300 liters per minute with a sampling
period of approximately 24 hours for a resulting total air volume of approximately 288 m ' 432
m^
The filters and sorbent cartridges will be solvent pre-cleaned and vacuum dried by the analytical
laboratory. The pre-cleaned filters and cartridges will be stored wrapped in aluminum foil (or
otherwise protected from light) in a Ziploc bag, and over wrapped with bubble wrap until they
are carefully installed on the sampler on site.
At the conclusion of each sampling period, the sample is recovered from the sampling train by
placing the filter inside the glass cartridge. The cartridge is then wrapped with aluminum foil
and it is placed back into its original shipping container, labeled, and transported to the analytical
laboratory for processing.
The amount of air sampled through the filters and cartridges will be recorded and the
filter/cartridge set placed in an appropriately labeled container and shipped to the laboratory for
analysis.
Equipment and supplies utilized with this sampling approach are as follows:
• High-Volume Sampler - General Metals Works model PS-1 (or performance equivalent).
Capable of pulling ambient air through the filter/sorbent cartridge at a flow rate of
approximately 200-300 1pm over a 24-hour sampling period.
• Sampling Module - Metal filter holder capable of holding a 102-mm circular particle filter
supported by a 16-mesh stainless-steel screen and attaching to a metal cylinder capable of
holding a 65-mm O.D. (60-mm I.D.) x 125-mm borosilicate glass sorbent cartridge
containing PUP and XAD-2® resin in the “sandwich” configuration. To ensure air-tight
seals, the filter holder is equipped with inert sealing gaskets placed on either side of the filter
and inert pliable gaskets placed at either end of the glass sorbent cartridge.
• High-Volume Sampler Calibrator - Capable of providing multi-point resistance for the high-
volume sampler.