Ambient Air Monitoring Program Section No. 10
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 59 of 116
Teflon end caps are replaced on the cartridge, and it is placed back into its original sealed and
labeled container, placed in an ice chest under ice (< 4°C), and transported to the analytical
laboratory for processing.
Equipment and supplies utilized with this sampling approach are as follows:
• Low-Volume, Continuous Flow Personal Sampling Pump - SKC Leland Legacy Personal
Monitor and Sampling Pump (or equivalent): 5 to 15 1pm.
• Battery Pack - Rechargeable lithium-ion battery pack (or equivalent) compatible with the
personal sampling pump.
• Sampling Cartridge - Constructed from a 20-mm (l.D.) by 10-cm borosilicate glass tube
drawn down to a 7-mm (O.D.) open connection for attachment to the pump by way of
flexible tubing.
• Polyurethane Foam (PUF) Plugs - Cut into a cylinder, 22-mm (l.D.) and 7.6-cm long, fitted
under slight compression inside the cartridge. Polyether type (density 0.0225 g/cm3). Plugs
should be slightly larger in diameter than the internal diameter of the cartridge. Pre-extracted
PUF plugs and glass cartridges may be obtained commercially.
• Teflon® End Caps - For sample cartridge. Must fit tightly to provide an adequate seal to
prevent pre- or post- sampling exposure to other potential sources of the target analytes.
• Glass Sample Cartridge - For sample collection. The analytical laboratory will insert the
PUF plug into the glass sample cartridge during pre-sampling preparation.
• Flexible Tubing - Used to connect the cartridge assembly to the sampling pump.
• Sample Cartridge Shipping Containers - The analytical laboratory will provide individually
labeled containers that will be used to transport the sample cartridges to the field location and
back to the laboratory for analysis. This will include aluminum foil wrapped around the
cartridge and a glass jar large enough to hold the cartridge.
• White Cotton Gloves - For handling the cartridges.
• Ice Chests - To hold samples at 4°C during shipment to the analytical laboratory.
• High Flow Rotometer - 2.0 to 20.0 1pm
L2005-346 Rev. 1
CITY_DEP_WORBY_CANNATA_SUBP_000114700
NYC-WTC_000428383