Ambient Air Monitoring Program Section No. 8
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 50 of 116
modifications to achieve quantitation limits that are significantly below the project Action
Limits. The laboratories selected will, at a minimum, meet the project quantitation limits
included in Table 8-1.
Laboratories will need to adjust all quantitation limits based on dilutions, sample volumes,
extract/digestate volumes, and cleanup procedures. In all cases, the adjusted quantitation limit
(or sample quantitation limit) must be below the project Action Limit. In establishing the
required quantitation limits for this program, these factors were considered in ensuring the
project Action Limits would be achieved.
Sensitivity will be maximized by following the EPA and NIOSH methods, utilizing
experienced/trained laboratory personnel, and conducting laboratory audits.
8.2.6 Completeness
Completeness is a measure of the amount of valid data obtained from a measurement system
compared to the amount that was expected to be obtained under normal conditions. “Normal
conditions” are defined as the conditions expected if the sampling plan was implemented as
planned.
Field Completeness Objectives
Field completeness is a measure of the amount of (1) valid measurements obtained from all the
measurements taken in the project and (2) valid samples collected. With the exception of the
real-time measurements, the field completeness objective will be a minimum of 80 percent. This
allows for the potential loss of samples due to sampling problems or media breakage during
transport. The completeness objective for the real-time measurements (e.g., PMio, PM2.5,
gaseous mercury) will be a minimum of 90 percent.
Laboratory Completeness Objectives
Laboratory completeness is a measure of the amount of valid measurements obtained from all
valid samples submitted to the laboratory. The laboratory completeness objective will be a
minimum of 90 percent. This allows for the potential loss of samples impossible to analyze due
to unforeseen interferences and rejected data following data validation.