Ambient Air Monitoring Program Section No. 8
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 48 of 116
Laboratory Accuracy Objectives
Laboratories assess the overall accuracy of their instruments and analytical methods
(independent of sample or matrix effects) through the measurement of “standards”, materials of
accepted reference value. Accuracy will vary from analysis to analysis because of individual
sample and matrix effects. In an individual analysis, accuracy will be measured in terms of
method blank results, processing blank results, the percent recovery (%R) of surrogate or internal
standard compounds, standard reference materials (SRMs) and/or laboratory control samples
(LCSs) and LCS Duplicates. This gives an indication of expected recovery for analytes tending
to behave chemically like the spiked or surrogate compounds and provides a measure of bias for
the parameter of interest. Accuracy control limits are provided in Tables 8-2 through 8-9. The
laboratory method blanks will indicate any adverse effects of sample contamination from an
outside source (i.e., sample preparation or sample analysis) and could result in a positive bias.
The frequency of surrogates or internal standards, SRMs, LCSs, LCS Duplicates are defined in
Tables 8-2 through 8-9. Laboratory accuracy will be improved by following the EPA and
NTOSH methods which include detailed requirements for each analysis, utilizing
experienced/trained laboratory personnel, ensuring the purity of all chemicals, and conducting
laboratory audits.
Accuracy of the silica analysis will be improved by an initial collection of a dust sample from the
site. The dust sample will be collected on each floor of the building, homogenized by the
laboratory, and analyzed for silica. This will be the site-specific standard used by the laboratory
for the remainder of the program and will serve the following two purposes:
• Will identify the phases of silica potentially present in the air
• Will show interferences that may require the use of secondary and/or tertiary peaks for silica
phase quantitation
8.2.3 Representativeness
Representativeness is a qualitative parameter which expresses the degree to which the data and
sampling design accurately and precisely represent a characteristic of a population, parameter
variations at a sampling point, a process condition, or an environmental condition within a
defined spatial and/or temporal boundary. Representativeness is a qualitative parameter which is
dependent upon the proper design of the sampling program and the laboratory quality control
program.
Measures to Ensure Representativeness of Field Data
Representativeness is dependent upon the proper design of the sampling program and will be
satisfied by ensuring that the Ambient Air Monitoring Program, referenced sampling
methodologies, and required QC procedures are followed and that proper sampling, sample
L2005-346 Rev. 1
C ITY_D E P_WO RB Y_C AN N ATA_SU B P_000114689
NYC-WTC_000428372