5.1 PMio Monitoring (“Real-Time”/Continuous)
5.1.1 Beta - Attenuation PMio Monitors (All Sites)
The monitors selected to continuously measure PMio are beta-attenuation monitors manufactured
by Met One Instraments, Inc. (Met One). The Met One E-BAM will be used for continuous
PMio measurements. The instrument operates on the principle of beta attenuation.
The E-BAM has not been officially designated by USEPA as either Reference or Equivalent
Method. However, the E-BAM design is descended directly from the BAM-1020 (which has
received USEPA’s designation as an Automated Equivalent Method - EQPM-0798-100),
modified to provide portable battery operation and produce measurements in real-time
(averaging times less than 1 hour). The accuracy and precision of the E-BAM are consistent
with USEPA requirements for Class III designation for PMio Class III equivalent method
instruments include any candidate instruments that cannot qualify as Class 1 or Class 11
instruments. These may either be filter-based integrated samples not meeting Class I or Class n
criteria, or filter or non-filter based continuous or semi-continuous samplers. Other methods
include all non-FRM or non-equivalent measurement methods capable of characterizing fine
particles that may not be or have not yet been classified as an equivalent method. Existing
manual and continuous analyzers in this category include the dichotomous sampler, IMPROVE
samplers, nephelomethers, beta attenuation monitors, and Tapered Element Oscillating
Microbalances (TEOMs). Such instruments are not precluded from becoming equivalent on a
site-specific, regional or national basis.
The beta attenuation process uses a small source of beta particles (carbon-14, 60 microcuries) is
coupled to a sensitive detector that counts the emitted beta particles. The dust particles are
collected on a filter tape that is placed between the beta source and the detector. Dust on the
filter will intercept some of the beta particles. The reduction of beta particles is proportional to
the amount of dust on the filter, which allows the mass of dust to be determined from the beta
particle counts. The dust mass is combined with the air volume collected during the filter
exposure time to determine the PM concentration.
The E-BAM monitors will be equipped with particle size selective inlets. The design of the
inlets is such that particles larger than the desired size range will be removed from the air flow,
based on the air flow rate. The units will be equipped with an inlet head to separate PMjo.
Sampling flow rate is critical to maintain the proper particle size cut points of the inlets. Flow
rates are maintained at 16.7 liters per minute (LPM) in the E-BAM monitors using an integral
flow meter, pressure sensor, and ambient temperature sensor on board each monitor.
The data from the E-BAM units will be recorded by digital data loggers using the analog signal
outputs of the monitors. The PMio data from the E-BAM monitors will be recorded as 5-minute,
hourly, and daily (midnight-to-midnight) averages.