• Personal Sampling Pump - The pump used to draw the air through the iodated
carbon trap.
• Flow Measurement Device - A device such as a rotameter, bubble meter, or gas
volume displacement-measuring device such as a “DryCal” flow meter. This
device should have a calibration traceable to a NIST certification.
• lodated Carbon Trap - The carbon trap provided by the subcontracted laboratory
that is used to absorb the mercury from the air that is passed through it.
10.2.9 PMio
PMio will be determined by two separate methods. At each location, PMio will be
determined by a non-reference method utilizing Met One’s E-BAM Mass Monitor, which
is a real-time monitor. Tisch TE-6070 High Volume MFC PM-10 MFC Monitor
Reference Method samplers will be used to collect PMio via the US EPA
standard reference method (40 CFR Part 50 Appendix J) at one location each day. A
Tisch TE-6070 sampler will be collocated along side the real-time PMio monitors as a QA
check. The Tisch TE-6070 sampler will rotate on a monthly basis through all real-time
PMio monitor locations for the duration of the monitoring program.
10.2.9.1 PMio-Reference Method via Tisch TE-6070 samplers
Particulate samples whose aerodynamic diameter is < 10 [J will be collected using Tisch
TE-6070 samplers operating in the PMio mode. All samples will be collected over an
estimated twenty-four hour sampling interval at a prescribed flow rate of 40 sefm. Pre
weighed Micro-Quartz filter media provided by the subcontracted laboratory will be
placed in the TE-6070 sampler prior to the start of each sampling event. The sampler
specification for the PMio sampler is contained in 40 CFR Part 50 Appendix J.
This document specifies the PMio sampling limits, protocol, and instrument
specifications. The TE-6070systQms perform all of the functions required or
recommended in the instrument specification part of the standard.
10.2.9.2 PMio - Continuous Monitoring of PMio via Met One E-BAM Samplers
The monitors selected to continuously measure PMio are beta-attenuation monitors
manufactured by Met One Instruments, Inc. (Met One). The E-BAM monitors will be
equipped with particle size selective inlets. The design of the inlets is such that particles
larger than the desired size range will be removed from the air flow at the prescribed air
flow rate. The units will be equipped with an inlet head to separate PMio. Sampling flow
rate is critical to maintain the proper particle size cut points of the inlets. Flow rates are
maintained at 16.7 liters per minute in the E-BAM monitors using an integral flow meter,
pressure sensor, and ambient temperature sensor on board each monitor.
10.2.10 PM2.5
PM2.5 will be determined by two separate methods. At each location, PM2.5 will be