SILICA. CRYSTALLINE, by XRD: METHOD 7500, Issue 3, dated 15 January 1998 - Page 4 of 8
watchglass with 2-propanol, collecting the washings in the beaker. 7. Place a silver filter in the filtration apparatus. Attach the funnel securely over the entire filter circumference. With no vacuum, pour 2 to 3 mL 2-propanol onto the filter. Pour the sample suspension from the beaker into the funnel and apply vacuum. During filtration, rinse the beaker several times and add rinsings to the funnel. 8. Control the filtration rate to keep the liquid level in the funnel near the top during rinsing. Do not wash the walls or add 2-propanol to the funnel when the liquid level is lower than 4 cm above the filter. Leave the vacuum on after filtration to produce a dry filter. Remove the filter with forceps and mount it in the XRD sample holder.
CALIBRATION AND QUALITY CONTROL:
9. Prepare and analyze standard filters. a. Prepare two suspensions of each analyte in 2-propanol by weighing 10 and 50 mg of the standard material to the nearest 0.01 mg. Quantitatively transfer each to a 1 -L glass-stoppered bottle using 1.00 L of 2-propanol. b. Suspend the powder in 2-propanol with an ultrasonic probe or bath for 20 min. Immediately move the flask to a magnetic stirrer with thermally insulated top and add a stirring bar. Cool to room temperature before withdrawing aliquots. c. Mount a silver filter on the filtration apparatus. Place several mL of 2-propanol on the filter. Turn off the stirrer and shake vigorously by hand. Immediately remove the lid and withdraw an aliquot from the center at half-height of the 10 mg/L or 50 mg/L suspension. Do not adjust the volume in the pipet by expelling part of the suspension. If more than the desired aliquot is withdrawn, discard the aliquot in a beaker, rinse and dry the pipet, and take a new aliquot. Transfer the aliquot from the pipet to the silver filter, keeping the tip of the pipet near the surface but not submerged in the delivered suspension. d. Rinse the pipet with several mL 2-propanol, draining the rinse into the funnel. Repeat the rinse several times. e. Apply vacuum and rapidly filter the suspension. Do not wash down the sides of the funnel after the deposit is in place since this will rearrange the material on the silver filter. Leave vacuum on until filter is dry. Transfer the silver filter to the diffractometer sample mount. Prepare working standard filters, in triplicate, at e.g., 20, 30, 50, 100, 200, and 500 pg. f. Analyze the working standards together with samples and blanks (step 12). The XRD intensities for the working standards (step 12.d) are designated“land are then normalized (step 12.e) to obtain Correct the intensifies of working standards >200 pg for matrix absorption (steps 12.f and 13). g. Prepare a calibration graph (3, vs pg of each standard). NOTE: Poor repeatability (>10% above 0.04 mg silica) at any given level indicates that new standards should be made. The data should lie along a straight line. A weighted least squares (W^ weighting) is preferable. h. Determine the slope, m, of the calibration graph in counts/pg. The intercept, b, on the abscissa should be within ± 5 pg of zero. NOTE: A large intercept indicates an error in determining the background, i.e., an incorrect baseline or interference by another phase. 10. NOTE: The following procedure for absorptioncorrection is not necessary in situations that have been previously documented as requiring no corrections. Select six silver membrane filters as media blanks randomly from the same box of filters to be used for depositing the samples. These will be used to test for sample self-absorption. Mount each of the media blanks on the filtration apparatus and apply vacuum to draw 5 to 10 mL 2-propanol through the filter. Remove, let dry, and mount on XRD holders. Determine the net normalized count for the silver peak, f,4g, for each media blank (step 12). Obtain an average value for the six media blanks.
MEASUREMENT:
11. Obtain a qualitative X-ray diffraction scan (e.g., 10 to 8028) of the area air sample (or bulk settled dust) to determine the presence of free silica polymorphs and interferences (see APPENDIX). The expected diffraction peaks are:
NIOSH Manual of Analytical Methods (NMAM), Fourth Edition
CITY_21MC1OO_O11O95724
NYC-WTC_000319956
OCR can misread numbers and units. Confirm readings against the page image before using them.
Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.
NYC-WTC_000338314
Analysis of crystalline silica forms (quartz, cristobalite, tridymite) by Method 7500.
5 shared non-person entities · 1 shared topics · similarity 0.986
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000338495
XRD-related documentation within WTC environmental testing files.
4 shared non-person entities · similarity 0.936
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000338486
MSD0700 silica related SOP for XRD analysis in WTC project files.
4 shared non-person entities · similarity 0.927
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000318556
Information sheet on asbestos design and analysis standard D5755, undated revision notes.
3 shared non-person entities · similarity 0.915
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000317863
Protocol and properties for asbestos and other fiber analysis using PCM; sampling methods listed.
3 shared non-person entities · similarity 0.902
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000363392
ATC TEM dust method practice document referencing AHERA/ASTM 5755.
4 shared non-person entities · similarity 0.921
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000281368
ICP element list and method for analysis; August 2004 issue reference.
7 shared non-person entities · 1 shared topics · similarity 0.903
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000281377
Table of elements used in ICP analysis; method and properties.
7 shared non-person entities · 1 shared topics · similarity 0.901
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000281386
Elements list for ICP analysis; includes aluminum, chromium, etc.
7 shared non-person entities · 1 shared topics · similarity 0.900
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000389978
Policy language on clean areas, HEPA filtration, and sampling expectations for asbestos work.
4 shared non-person entities · similarity 0.899
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000435596
Lab report describing asbestos and other fibers by PCM with guidelines and CAS numbers.
4 shared non-person entities · similarity 0.897
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000433519
PCM/TEM asbestos fiber analysis parameters and reference values.
4 shared non-person entities · similarity 0.897
machine-extracted · confidence unavailable · Check page ↗Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.
NYC-WTC_000338314
Analysis of crystalline silica forms (quartz, cristobalite, tridymite) by Method 7500.
Similarity score 0.999 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000338314
Analysis of crystalline silica forms (quartz, cristobalite, tridymite) by Method 7500.
Similarity score 0.951 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000338495
XRD-related documentation within WTC environmental testing files.
Similarity score 0.942 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000338495
XRD-related documentation within WTC environmental testing files.
Similarity score 0.940 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000338486
MSD0700 silica related SOP for XRD analysis in WTC project files.
Similarity score 0.937 (not a probability)
machine-extracted · confidence unavailable · Check page ↗NYC-WTC_000281386
Elements list for ICP analysis; includes aluminum, chromium, etc.
Similarity score 0.931 (not a probability)
machine-extracted · confidence unavailable · Check page ↗Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.
NYC-WTC_000338314
Analysis of crystalline silica forms (quartz, cristobalite, tridymite) by Method 7500.
Similar indexed content
NYC-WTC_000338495
XRD-related documentation within WTC environmental testing files.
Similar indexed content
NYC-WTC_000338486
MSD0700 silica related SOP for XRD analysis in WTC project files.
Similar indexed content
NYC-WTC_000363392
ATC TEM dust method practice document referencing AHERA/ASTM 5755.
Similar indexed content
NYC-WTC_000318556
Information sheet on asbestos design and analysis standard D5755, undated revision notes.
Similar indexed content