Silica, crystalline by XRD method details

Machine-extracted title · confidence 42%

Laboratory method sheet for crystalline silica analysis by XRD (Quartz, cristobalite, tridymite).

NYC-WTC_000319953–000319960
Page image
Page image not rendered yet.
The pipeline renders page images incrementally; check back or read the OCR text alongside.
OCR text

OCR status: ok · source: pdftotext

NYC 9/11 Public Portal Document

SILICA, CRYSTALLINE, by XRD: METHOD 7500, Issue 3, dated 15 January 1998 - Page 5 of 8

Peak (2-Theta Degrees) Mineral Primary Secondary Tertiary Quartz 26.66 20.85 50.16 Cristobalite 21.93 36.11 31.46 Tridymite 21.62 20.50 23.28 Silver 38.12 44.28 77.47 NOTE: There is an alternative to scanning an area air sample, settled dust sample, or ground bulk sample to prove lack of contamination. A slow scan of the three main peaks of quartz (also cristobalite andtridymite if their absence has not been previously confirmed) on a personal air sample, and verification that their intensity ratios are within 15®/o of pure quartz, is sufficient evidence that other materials are not interfering in the silica determination. 12. Perform the following for each sample, working standard, and blank filter; a. Mount the reference specimen. Determine the net intensity,,lof the reference specimen before and after each filter is scanned. Use a diffraction peak of high intensity that can be rapidly but reproducibly <0.01) measured. b. Mount the sample, working standard, oiblank filter. Measure the diffraction peak area for each silica polymorph. Scan times must be long, e.g., 15 min (longer scan times will lower the limit of detection). c. Measure the background on each side of the peak for one-half the time used for peak scanning. The sum of these two counts is the average background. Determine the position of the background for each sample. d. Calculate the net intensity, (, (the difference between the peak integrated count and the total background count). e. Calculate and record the normalized intensity,j for each peak:

NOTE: Select a convenient normalization scale factor, N, which is approximately equivalent to the net count for the reference specimen peak, and use this value of N for all analyses. Normalizing to the reference specimen intensity compensates for long-term drift in X-ray tube intensity. If intensity measurements are stable, the reference specimen may be run less frequently and the net intensities should be normalized to the most recently-measured reference intensity. f. Determine the normalized count, Kg, of an interference-free silver peak on the sample filter following the same procedure. Use a short scan time for the silver peak (e.g., 5% of scan time for analyte peaks) throughout the method. g. Field blanks may be analyzed by scanning the 2-theta range used for the analyte and silver peaks to verify that contamination of the filters has not occurred.The analyte peak should be absent. The normalized intensity of the silver peak should match that of the media blank. Each laboratory should determine the specifics of field blank use for its application. When contamination does occur, the reason should be investigated and appropriate action taken. In practice, contamination of field blanks is extremely rare and usually is not consistent across filters. The analysis of blanks may be abbreviated if experience indicates that contamination is not likely with current field and laboratory operations; however, occasional confirmation of non-contamination is prudent.

CALCULATIONS:

13. Calculate the concentration of crystalline silica, C (mg/^, in the air volume sampled, V (L): ■x • f(t) - b C ---------------- , mg/m^ m V

lx normalized intensity for sample peak b intercept of calibration graph (J vs. W)

NIOSH Manual of Analytical Methods (NMAM), Fourth EdiUon

CITY_21MC100_0 11095725

NYC-WTC_0003 19957

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000319957Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

Compare copies and versions →