Silica, crystalline by XRD method details

Machine-extracted title · confidence 42%

Laboratory method sheet for crystalline silica analysis by XRD (Quartz, cristobalite, tridymite).

NYC-WTC_000319953–000319960
Page image
Page image not rendered yet.
The pipeline renders page images incrementally; check back or read the OCR text alongside.
OCR text

OCR status: ok · source: pdftotext

NYC 9/11 Public Portal Document

SILICA, CRYSTALLINE, by XRD: METHOD 7500, Issue 3, dated 15 January 1998 - Page 3 of 8

SAMPLING:

1. Calibrate each personal sampling pump with a representative sampler in line. 2. Sample at 1.7 ± 5% L/min with nylon cyclone, 2..2 + 5% L/min with HD cyclone, ro 2.5 ± 5% Al cyclone for a total sample size of 400 to 1000 L. Do not exceed 2 mg dust loading on the filter. NOTE: Do not allow the sampler assembly to be inverted at any time when using a cyclone. Turning the cyclone to anythingmore than a horizontal orientation may deposit oversized material from the cyclone body onto the filter. 3. Take an area air sample or collect a settled dust sample, if dust in the work environment has not been previously characterized.

SAMPLE PREPARATION:

4. Samples may be characterized by one of the following methods, as appropriate. a. Interference check. Prepare area dust sample or settled dust bulk sample for XRD analysis by mounting the collection sample directly on an XRD sample holder, or by depositing or redepositing the dust on another filter for mounting, or by packing an XRD powder holder. Proceed to step 11. b. Qualitative Analysis. Prepare the area air sample or settled dust sample for qualitative analysis by grinding and/or wet sieving to best match the airborne dust particle size. Wet sieve with a 10- pm sieve. 2-propanol, and an ultrasonic bath [11], followed by evaporation of excess alcohol, drying in an oven for 2 hours, and overnight storage in a desiccator. Deposit the end product on a filter (steps 7-8) or pack in a conventional XRD powder holder. NOTE 1: For quantitative determination of % SiQ. weigh out, in triplicate, 2 mg sieved dust, transfer to a 50-mL beaker, add 10 mL 2-propanol, and continue with step 6. NOTE 2: In a bulk sample, if there is an interfering compound(s) that renders the identification and quantitation of quartz very difficult, the sample will need to be carefully treated in hot phosphoric acid [12] to dissolve the interfering compound(s) and avoid the loss of quartz. This treatment can be used to dissolve several 50-mg sample aliquots in order to concentrate the quartz content for the purpose of lowering the LOD. 5. Use one of the following methods to prepare filter samples and blanks: a. Low Temperature Ashing: Place the filters in 50-mL beakers within the low temperature asher so that the sample exposure to the plasma is optimized. Ash according to manufacturer's instructions. After ashing, carefully add 15 mL 2-propanol to each beaker; or b. Muffle Furnace Ashing: 1. If the samples contain a significant amount of calcite (>20% of total dust loading), silica may be lost due to formation of CaSiQ. Remove the calcite by the following procedure: Place a 0.5-pm, 25-mm PVC filter in the filtration apparatus and clamp the filter funnel over it. Remove the sample filter from the cassette, fold, and drop it on the 25-mm filter. Add 10 mL 25% v/v HCl and 5 mL 2-propanol to the filter funnel and allow to sit for 5 min. Apply vacuum and slowly aspirate the acid and alcohol in the funnel, washing with three successive 10-mL portions of distilled water. Release the vacuum. Carry both filters through the ashing step together. 2. Place the filter samples in porcelain crucibles, loosely cover and ash imuffle furnace for 2 h at 600 °C (800 °C if graphite is present). Add several mL 2-propanol to the ash, scrape the crucible with a glass rod to loosen all particles and transfer the residue to a 50-mL beaker. Wash the crucible several more times and add wash to beaker. Add 2-propanol to the beaker to bring the volume to about 15 mL; or c. Filter Dissolution: Using forceps and a spatula, remove the filter from the cassette, fold the filter three times, and place in the bottom of a 40- or 50-mL centrifuge tube. Add 10 mL THF and allow to stand for at least 5 min. Cap the centrifuge tube with aluminum foil to prevent contamination. Gently agitate the centrifuge tube by hand or with a'ortex mixer making sure the THF does not go near the top of the tube. Place the tube in an ultrasonic bath (water level 2.5 cm from top) for at least 10 min. (The filter should be totally dissolved.) Just prior to filtering, agitate the sample for 10 to 20 sec on a vortex mixer. Continue with step 6, substituting THF for 2-propanol and centrifuge tube for beaker. 6. Cover the beaker with a watchglass and agitate in an ultrasonic bath for at least 3 min. Observe the suspension to make sure that the agglomerated particles are broken up. Wash the underside of the

NIOSH Manual of Analytical Methods (NMAM), Fourth Edition

CITY_21MC100_011095723

NYC-WTC_000319955

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000319955Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

Compare copies and versions →