Bulk asbestos analysis method update, lab manual

Machine-extracted title · confidence 48%

EPA/DOE manual describing identification of asbestos in non-friable bulk materials.

NYC-WTC_000319762–000319785
Page image
Page image not rendered yet.
The pipeline renders page images incrementally; check back or read the OCR text alongside.
OCR text

OCR status: ok · source: pdftotext

NYC 9/11 Public Portal Document

FROM US EPA/OERR (TUE) 1 2. 1 1 ' 01 9:17/ST. 9 : 1 1/NO. 486 1 202483 P 10

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL

SUBJECT DATE PAGE DTEM NO.

Transmission Electron Microscope Method for 3/1/97 9 of 34 198.4 Identifying and Quantitating Asbestos in Non- Friable Organically Bound Bulk Samples

3.19.7.2. Low-magnification objective (3.2 to 10X). 3.19.7.3. High-magnification, dry objective (30 to SOX). 3.19.8. Eyepiece(s) of at least 8X magnification containing a fixed cross-hair. 3.19.8.1. Chalkley point-count reticle (optional). 3.19.9. Focusable condenser with centerable iris diaphragm capable of completely eclipsing the back-focal-plane image of the central stop. 3.20. Beam balance with readability of 1 mg or less, (e.g., Fisher Model 711). 3.21. Transmission electron microscope that must meet the specifications outlined in the AH ERA method for airborne asbestos analysis (EPA40 CFR 763 - Appendix A to Subpart 3, October 30. 1987 - Non-mandatory Method). 3.21.1 Specimen holder with at minimum a single-tilt capability through at least ±25°. 3.21.2 Calibration lines on the TEM's phosphor screen must be drawn in at least two mutually perpendicular directions or drawn as calibrated-diameter circles. A single calibration line is acceptable only if the specimen holder or image is capable of continuous rotation during analysis without affecting magnification calibration. 3.22. Mocropipette capable of dispensing 2-5 pL aliquots. 3.23. Disposable plastic beakers, approximately 25-50 mL capacity. 3.24. Analysis sheet (see Example 1 at end of Item 198.4) with space for the following entries; 3.24.1. Analyst's signature. 3.24.2. Date of analysis. 3.24.3. Sample identification number.

CITY_21MC1OO_OO3O5O1O5

NYC-WTC_000319770

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000319770Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

NYC-WTC_000342629

Summary of bulk and TEM sampling with lab analysis and quality control.

Similar indexed content

Compare copies and versions →