(TUE) 1 2. 1 1’ 01 9:27/ST. 9 : 1 1/NO. 4 861 20248 3 P 25 FROM US EPA/OERR

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL

SUBJECT DATE PAGE STEM NO.

Transmission Electron Microscope Method for 3/1/97 24 of 34 198.4 Identifying and Quantitating Asbestos in Non- Friable Organically Bound Bulk Samples

8.3. Contamination Control. 8.3.1. Prevention. The laboratory must detail its methods for preventing cross " cohtamihation of eq uipTneht, supplies and reagents. Much of this will be careful cleaning of work area, equipment and supplies. Intensity and frequency of this effort should be based on experience gained through any contamination detected as described in 8.3.2. 8.3.2. Monitoring. The laboratory must have a routine procedure for monitoring contamination of laboratory equipment, supplies and work stations and for resolving contamination problems when discovered. At least one asbestos-free NOB must be prepared and analyzed with every 20 samples analyzed. This non-ACM must go through the full preparation and analysis regimen for the type of analysis being performed. If asbestos is detected, the source of contamination must be traced and the problem resolved to prevent recurrence. Any of the previous samples that may have had results affected by the contamination must be reanalyzed and the client notified of any revisions to reported values. Detailed records of monitoring must be maintained. 8.4. Calibration. Written records for calibration of the following equipment and supplies must be kept. All calibrations listed below (unless otherwise noted) must be performed under the same analytical conditions used for routine asbestos analysis and must be recorded in a bound notebook and include date and analyst's signature. Frequencies stated below may be reduced to "before next use" if no samples are analyzed after the last calibration period has expired. Likewise, frequencies should be increased following non-routine maintenance or unacceptable calibration performance.

CITY_21MC100_003050120

NYC-WTC_000319785

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000319785Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

NYC-WTC_000342629

Summary of bulk and TEM sampling with lab analysis and quality control.

Similar indexed content

Compare copies and versions →