(TUE) 12. 11’01 9 : 21/ST. 9:1 1/NO. 4861202483 P 16 FROM US EPAZOEHR

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL

SUBJECT DATE PAGE ITEM NO.

Transmission Electron Microscope Method for 3/1/97 15 of 34 198.4 Identifying and Quantitating Asbestos in Non- Friable Organically Bound Bulk Samples

concentrations will exceed 10%. For these high-concentration ACMs, estimations can be based on a single grid. In making this estimation, the analyst should keep in mind the relative thickness of the different materials on the grid. The amount of asbestos is then calculated by; % Asbestos = (PAM/OM)X(AOS) where. PAM = Mass of residue after furnace and acid treatments (mg) OM = Mass of original subsample (mg) AOS = Mean percentage of asbestos (versus inorganic residue) in both grid preparations 5.3. Polarized-Liqht Microscope Analysis. If TEM analysis yields asbestos concentrations less than 1% and PLM analysis has not been performed in accordance with Section 5.2.2 of Item 198.1, two coverslip preparations must be made according to Section 4.5 and examined by PLM. The entire areas under both coverslips must be examined for traces of asbestos. If any asbestos is detected by PLM, two additional coverslip preparations must be made and all four must be quantitated in accordance with Section 5.2.2 of ELAP Item 198,1. 5.4. Analytical Records. Detailed records must be kept of all phases of the analysis. An analysis sheet that includes all the data required in Section 3.25 must be filled out completely, signed and dated by analyst. 6. Test Reports. Reports to clients must include at least the following; 6.1. Cli^. Identify name and address. 6.2. Sample Identity. The identification number (Section 8.1.5.1.3) assigned by the laboratory must be clearly cross-referenced to information provided by the

CITY_21MC1001003050111

NYC-WTC_000319776

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000319776Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

NYC-WTC_000342629

Summary of bulk and TEM sampling with lab analysis and quality control.

Similar indexed content

Compare copies and versions →