ASBESTOS and OTHER FIBERS by PCM: METHOD 7400, Issue 2, dated 15 August 1994 - Page 2 of 15

REAGENTS: EQUIPMENT:

1. Acetone,* reagent grade. 1. Sampler: field monitor, 25-mm, three-piece 2. Triacetin (glycerol triacetate), reagent grade. cassette with ca. 50-mm electrically conductive extension cowl and cellulose ester filter, 0.45- to 1.2-pm pore size, and backup pad. NOTE 1: Analyze representative filters for * See SPECIAL PRECAUTIONS. fiber background before use to check for clarity and background. Discard the filter lot if mean is >5 fibers per 100 graticule fields. These are defined as laboratory blanks. Manufacturer-provided quality assurance checks on filter blanks are normally adequate as long as field blanks are analyzed as described below. NOTE 2: The electrically conductive extension cowl reduces electrostatic effects. Ground the cowl when possible during sampling. NOTE 3: Use 0.8-pm pore size filters for personal sampling. The 0.45-pm filters are recommended for sampling when performing TEM analysis on the same samples. However, their higher pressure drop precludes their use with personal sampling pumps. NOTE 4: Other cassettes have been proposed that exhibit improved uniformity of fiber deposit on the filter surface, e.g., bellmouthed sampler (Envirometrics, Charleston, SC). These may be used if shown to give measured concentrations equivalent to sampler indicated above for the application. 2. Personal sampling pump, battery or line- powered vacuum, of sufficient capacity to meet flow-rate requirements (see step 4 for flow rate), with flexible connecting tubing. 3. Wire, multi-stranded, 22-gauge; 1", hose clamp to attach wire to cassette. 4. Tape, shrink- or adhesive-. 5. Slides, glass, frosted-end, pre-cleaned, 25 x 75-mm. 6. Cover slips, 22- x 22-mm, No. 1-1/2, unless otherwise specified by microscope manufacturer. 7. Lacquer or nail polish. 8. Knife, #10 surgical steel, curved blade. 9. Tweezers.

NIOSH Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94

NYC-WTC_000149406

OCR can misread numbers and units. Confirm readings against the page image before using them.

NYC-WTC_000149406Source: NYC Law Department, mirrored locally

Related records

Browse subjects →

Ranked by indexed similarity. Reasons describe shared subjects and filing context; check the source records.

Same-box records

DEP WTC General Correspondence Form · Box DEP Box 10

NYC-WTC_000149798

Generic form or header page from the Department of Environmental Protection regarding general World Trade Center notices and requests.

19 shared non-person entities · Same agency, volume and box · similarity 0.961

machine-extracted · confidence unavailable · Check page ↗

Filed elsewhere

More like this page

Similar subjects in other documents. Similarity does not establish the same event, measurement or conclusion.

Page similarity is temporarily unavailable or this page has no indexed vector.

Buildings in this record

Machine-extracted building matches; verify each source. Buildings, never households.

Others also read

Readers open these alongside this record — through an Ask answer, the same folder, or indexed similarity. Not a claim they are about the same event.

Compare copies and versions →