Part V - Quality Assurance Project Plan
133-135 Greenwich Street/21-23 Thames Street
Airtek Project 05-0701 - November 4, 2005
10.3.5 Breakthrough Checks
Separate analyses of both front and back half portions of carbon traps for mercury will
be performed during the background phase to allow for an assessment of analyte
breakthrough. The breakthrough assessment will be performed on all stations during the
background phase. Breakthrough on the back half of the trap must be < 2% of the front
half of the trap or < 5 ng per trap. If breakthrough criteria are consistently met during the
background phase, subsequent measurement will be performed with the front and back
halves of the carbon traps combined for analyses.
10.3.6 Media Certification Checks
Media for all analyses collected a on time-integrated basis will be certified clean from
the laboratories. The certifications will be performed as batch checks and results will be
stored in the project files.
11.0 ANALYTICAL REQUIREMENTS
11.1 Analytical Methods
This section of the QAPP describes the analytical techniques that will be used by the
fixed laboratories to generate definitive data for the project. It documents the fixed
laboratory analytical methods that will be used to meet measurement performance criteria
and achieve the project-required quantitation limits for all contaminants of concern in the
specific matrices as identified on Tables 8-la-c.
11.1.1 Fixed Laboratory Analytical Methods and SOPs
Table 11-1 details the analytical methods that will be used in this investigation
11.1.2 Fixed Laboratory Analytical Method/SOP Modifications
The cited methods will be followed as written with the exceptions summarized below.
These modifications do not adversely affect the quality or usability of the data as
sufficient quality control analyses will be performed demonstrating adequate method
performance with these modifications.
11.1.2.1 PAH Extraction by EPA Metlwd TO-13A and Analysis by SW-846 Method
8270C
The analysis of PAHs using SW-846 Method 8270C will be modified to include the use
of selective ion monitoring (SIM). Modifications to the sample preparation procedure in
EPA Method TO-13 are listed below.
68
CITY_21MC100_003237472