cassette. At the end of the sampling period, the top cover and small end caps are replaced.
Samples are transported upright to the laboratory in a rigid container with packing
material to prevent jostling or damage. To avoid electrostatic forces from causing fiber
loss from the sampling filter, untreated polystyrene foam will not be used for any of the
shipping container materials. Equipment and supplies utilized with this sampling
approach are as follows:
• Area Sampling Pump - 5 to 15 1pm.
• Filters/TEM & PCMe analysis - Mixed cellulose ester membrane filters, 0.45-Lini
pore size, with a static-conductive cowl, supported by a cassette filter holder,
suitable for connection with the personal monitoring pump system.
• High Flow Rotometer - 2.0 to 20.0 1pm
10.2.4 Respirable Crystalline Silica
Respirable crystalline sampling will be performed according to NIOSH Method 0600
- Particulates Not Otherwise Regulated, Respirable following the recommendations for
personal sampling. Samples will be collected at a rate of approximately two liters per
minute (2 1pm) for a minimum of eight (8) hours for a resulting total air volume of
approximately 1.0 m3. To ensure optimum accuracy, the sampling rate will be adjusted,
as needed, not to exceed 2 mg dust loading on the filter. Filters are pre-weighed to a
constant weight in the controlled weighing area. If static electricity is evident (e.g., filter
not releasing easily from the forceps or it attracts the balance pan), the filter should be
passed over an anti-static radiation source. Filters are assembled in the filter cassettes and
the cassettes closed firmly by placing a plug at each opening of the cassette to
prevent leakage around the filter. For at least 2 hours prior to sampling, the filters are
equilibrated in either an environmentally controlled weighing area or chamber. Pump
flow rates of each personal sampling pump, with a representative sampler in line
and cyclone grit pot in place, will be measured before and after sample collection. The
pump will be run for 10 minutes prior to checking the flow rate. A suitable rotometer (or
bubble meter) will be connected to the filter inlet, the flow screw adjusted to the desired
sampling value, and the flow rate observed on the rotometer (or bubble meter) will be
verified as constant for a minimum of 20 seconds. Flow rates will be recorded on Field
Sampling Data Sheets or an equivalent field logbook. The cyclone’s grit cap is removed
prior to use and the interior inspected. If the inside is visibly scored, replace the cyclone
because the dust separation characteristics may be altered. Clean the interior of the
cyclone to prevent re-entrainment of large particles. Assemble the sampler head checking
for alignment of the filter holder and cyclone in the sampling head to prevent leakage.
Care must be taken to prevent the sampler assembly from inverting at any
time. Immediately prior to collecting a sample, the sampler will be fastened to a location
(i.e., on a tripod) near the individual’s breathing zone, the top cover of the filter cassette
cowl extension (“open face”) will be removed and positioned “face down” in the
sampler. The joint between the extender and monitor body will be wrapped with tape to
hold the cassette securely in place and to provide a surface to identify the cassette. By