Part V - Quality Assurance Project Plan
133-135 Greenwich Streel/21-23 Thames Street
Airtek Project 05-0701 - Rev 1 - Eebruary 1, 2006
Recalibrate System GCPEST011799
Initial cal. (5 As needed RSDs all <20%
GC Pesticides/ stds) Cont. Rerun all affected Revision 1.1
Every 10 < 15 % Difference
Herbicides and calibration samples under good GCHERB011999
samples
PCBs (midpoint)
Not detected cev Revision 1.2
one per Within Lab control limits
method blank Reextract, rerun
batch GCPCB011799
surrogate stds. Within lab control limits Re extract samples
All Revision 1.2
MS/MSD & with both surrogates
Once per out
LCS
20 samples
Reextract or
reanalyze affected
samples where LCS
is out.
GCMSSVOC011700
DETPP Every 12 SW846-8270C criteria Retune, rerun for
GC/MS Semi Tuning hours DETPP criteria Revision 1.4
CCCs <30%RSD, SPCCs
Volatiles Initial cal. (5 As needed > 0.05 Rf; RSD avg.< 15% Recalibrate
stds) 1/20 per No targets found Re extract mblk
method blank matrix CCCs @ <25%D; SPCCs> Rerun or recal/rerun
Cont. every 12 0.05Rf all afftected samples
calibration hours Within Lab control limits Re-extract
Surrogate stds. ALL Within Lab control limits Reclibrate/ rerun all
Laboratory 1/20 per samples
Control matrix -50 to +100% of 12 hour Reextract & rerun
MS/MSD 1/20 per cal chk std. Rerun affected
Internal matrix samples, dilute if
Standards ALL matrix is issue
Hgl20998
Initial cal. (5 As needed 0.995 Corr. Coeff. Recalibrate system
Mercury/eVA stds) Revision 1.1
Every 20 Not detected Rerun affected
A metliod blank samples samples
+/- 20%
Cont. Every 10 Spike -I-/- 25%, Dup +/- Rerun affected
calibration samples 20% at levels >5x samples
Spike and Dup 1/20 Reporting Limit Rerun spike/dup
samples Mfgs. Specs. unless proven matrix
interference
LCS
1/20 Perun all affected
samples samples
Recalibrate ICP 1CP031195
Initial As needed +/- 10%
system Revision 1.2
ICP metals calibration Every 10 +/- 10%
Rerun affected
Cont. samples Not detected (<IDL) samples
calibration Every 10
Rerun or qualify
Calibration samples
affected samples
blanks
Calibrate with On use +- 0.05 units Recalibrate, if still > SOPWCpH
Corrosivity/ pH=7, and 0.05 replace probe 010596 Rev.1.6
pH=4 buffers
pH