Ambient Air Monitoring Program Section No. 13
Quality Assurance Project Plan Revision No.: 0
130 Liberty Street - NY, NY Date: August 2005
Lower Manhattan Development Corporation Page 94 of 116
13.3.2 Analytical Laboratory Supplies/Consumables
Critical supplies and sample containers will be inspected in the following manner.
Critical Supplies and Inspection Requirements Responsible
Consumables and Acceptance Criteria Individual
Sample bottles, media Visually inspected upon receipt for cracks, breakage, cleanliness. Sample
Must be accompanied by certificate of analysis. Custodian
Chemicals and reagents Visually inspected for proper labeling, expiration dates, Laboratory
appropriate grade. Record lot numbers of reagents used for Analyst
standard preparation.
Supplies and consumables not meeting acceptance criteria will initiate the appropriate corrective
action. Corrective measures may include notification of vendor and subsequent replacement of
defective or inappropriate materials. All actions will be documented in the project files.
The use of materials of known purity and quality will be utilized for the calibration of all
instruments as part of this project. The laboratories will carefully monitor the use of all
laboratory materials including solutions, standards and reagents through well documented
procedures.
All solid chemicals and acids/bases used by the laboratories will be reagent grade or better. All
gases will be high purity or better. All standards or standard solutions will be obtained from U.S.
Environmental Protection Agency certified commercial sources.
All materials including standards or standard solutions will be dated upon receipt, and will be
identified by material name, lot number, purity or concentration, supplier, receipt/preparation
date, recipient/preparer’s name, and expiration date.
Standards or standard solution concentrations will be validated prior to use. This validation may
be re standardization for acids and bases, response factor comparison, standard curve response,
comparison to other standards made at a different time and/or by a different analyst. All
standards and standard materials will be checked for signs of deterioration including unusual
volume changes (solvent loss), discoloration, formation of precipitates or changes in analyte
response. All standards and standard solutions will be properly stored and handled and will be
labeled with all appropriate information including compound/solution name, concentration,
solvent, expiration date, preparation date, and the initials of the preparer.
All solvent materials or materials used as part of a given procedure will also be checked. Each
new lot of solvent will be analyzed to ensure the absence of interference.