io Ambient Air Monitoring Program
Quality Assurance Project Plan
130 Liberty Street - NY, NY
Section No. 13
Revision No.: 0
Date: August 2005
co
I Lower Manhattan Development Corporation Page 92 of 116 ■D
c
o
o CT
o o'
to ■D
00 o
-IS-
3.
CD 03
l ablc 13-4. Summary of Calibration Procedures for Fixed Laboratory AnaIvscs o
Paranieter/Instrunient o
Frequency of Calibration Acceptance Criteria Corrective Action o
c
Asbestos/TEM Alignment of TEM: performed daily NA Perform necessary equipment 3
maintenance and check calibration CD
O
standards.
Magnification: monthly 2x SD must be < 5% cumulative
mean
Camera Constants: monthly 2x SD must be < 5% cumulative
mean___________________________
Chrystotile Beam Dose: quarterly Fibrils must be visible for minimum
of 15 seconds____________________
Spot Diameter: quarterly Variation of spot diameters must be <
25% of the mean
EDXA Resolution: quarterly Mn K., peak has resolution < 175 eV Perform necessary equipment
at full width half maximum_________ maintenance and check calibration
Plasma Asher: quarterly Used to calculate time needed to standards.
remove 10% of collapsed mixed
cellulose ester filter_______________
Grid Opening Measurement: Variation of grid openings must not
O performed on 2% of grid lot to be > 5% of the mean
determine average grid opening in
mm^
O
m Silica/XRD Standard: daily analysis___________ 80-120% of tme value__________ Perform necessar y equipment
Initial Calibration: whenever there is Acceptable daily standard analysis maintenance and check calibration
a change in hardware or when a new results calculated using this curve; standards.
o set of calibration reference standards intercept value less than 5 pg for
J]
□0 are prepared primary peak curves
PMio Balance Calibration: daily before use ±3 pg Perform necessary equipment
s
> and every 10 samples (200.000 mg) maintenance and check calibration
Calibration Verification: daily before ±3 pg standards.
use (100.000 mg)
c
□0
o
o
o
L2005-346 Rev. 1