o Ambient Air Monitoring Program Section No. 8 o
co
g Quality Assurance Project Plan Revision No.: 0
I
o 130 Liberty Street - NY, NY Date: August 2005
“0
o Lower Manhattan Development Corporation Page 40 of 116 c
o CT
o
o'
ro "0
oo o
Ca5
o>
4^
Table 8-2. Measurement Performance Criteria Table - Metals by ICP/MS 03
QC Sample or Activity Frequency Measurement Performance Criteria Corrective Action DQI O
Field Duplicates 1/week *RPD < 40 when positive results for both Assess laboratory precision, qualify Precision-Overall o
o
samples arc > 5x QL data, and/or resample data. c
3
*No situation where one result is detected o
3
at > 5x QL and other result is not detected
Laboratory Duplicates l/prep batch RPD < 20 if results are > 5x QL Qualify data. Precision-Laboratory
Laboratory Control Sample 1/prcp batch Percent recoveries 75-125% Determine cause of problem, Accuracy/Bias
reanalyze, and/or qualify data.
Laboratory Control Sample l/prep batch Percent recoveries 75-125%; RPD < 20 Determine cause of problem, Accuracy/Bias and
Duplicate reanalyze, and/or qualify data. Precision
Serial Dilution Analysis 1/batch + 10% of original result Qualify data. Accuracy/Bias
Interference Check Sample 1/8 hours Percent recoveries 80-120% Recalibrate and reanalyze and/or Accuracy/Bias
qualify data.
Calibration Blanks 1/10 samples Absolute value of target metal must be Reclean, restest, reanalyze, and/or Accuracy/Bias -
<QL qualify data. Contamination
Preparation Blanks l/prep batch Absolute value of target metal must be Reclean, restest, reanalyze, and/or Accuracy/Bias -
<QL qualify data. Contamination
Field/Trip Blanks 1/day Absolute value of target metal must be Reclean, restest, reanalyze, and/or Accuracy/Bias -
<QL qualify data. Contamination
O
Quartz Fiber Filter Media 1/batch Absolute value of target metal must be Obtain new lot of media. Accuracy/Bias -
Cleaning Verification Check <QL Contamination
Internal Standards Every sample, Samples: 30-130% of IS calibration blank Dilute sample 5x, add IS and Accuracy/Bias
O blank, QC Calibration Standards: 80-120% of IS in reanalyze, and/or qualify data.
m
calibration blank
Instrument Tune Daily, prior to Mass resolution <1.0 amu at 10% peak Retune instrument, repeat tune Accuracy/Bias
o calibration and height and mass calibration + 0.1 amu solution and analysis.
J] sample analysis from the expected value
□0
Data Completeness Check NA Field 80%, Laboratory 95% NA Data Completeness