Part II - Environmental Community Monitoring Plan
133-135 Greenwich & 21-23 Thames Streets
Airtek Project 05-0701 - Rev 2 September 16, 2005
4/yr (manual)
External Leak Check 4/yr < 80 mL/min not described
Internal Leak Check 4/yr < 80 mL/min not described
Temperature Audit 4/yr -I- 2C not described not described
Pressure Audit 4/yr (?) ±10 mm not described
Balance Audit 1/yr Manufacturers specs not described
Precision
Collocated samples every 6 days for 25% of sites CV < 10%' Part 58, App.A, Sec
Single analyzer 1/3 mo. CV < 10% 3.5/5.5
Single Analyzer 1/yr CV < 10% not described
Reporting Qrg. 1/ 3 mo. CV < 10% not described
not described
Calibration & Check
Standards
Flow Rate Transfer Std. 1/yr +2% of NIST-traceable Std. Part 50, App.L Sec 9.1
and 9.2
Field Thermometer 1/yr + 0.1 C resolution not described
+ 0.5 C accuracy not described
Field Barometer 1/yr + 1 mm Hg resolution not described
+ 5 mm Hg accuracy not described
Working Mass Stds. 3-6 mo. 0.025 mg not described
Primary Mass Stds. 1/yr 0.025 mg not described
Table A-1 - Measurement Quality Objectives - Parameter PMIO (Dichotomous Sampler)
Requirement Frequency Acceptance Criteria Reference Information/Action
Reporting Units All data Pg/md 40 CFR, Pt 50.7
Filter Checks
Visual defect check Filter All filters See reference Vol II, MS 2.10.4 Discard any defective
Integrity filters
Collection efficiency Purchase 99% 40 CFR, Pt 50,
Integrity Specification + 5 pg/m3 App M, S 7.2 As measure by DOP test
Alkalinity <25.0 (ASTM-2988). Reject
Filter Conditioning All filters microequivalents/gram shipment.
Equilibration time 40 CFR, Pt 50, Following 2 months storage
Temperature range at least 24 hours App M, S 9.3 at ambient temp and
Temperature control 15 to 30° C 4 CFR, Pt 50, App relative humidity.
Humidity range + 3C M, S 7.4 Repeat equilibration
Humidity control 20 to 45 % relative humidity Keep thermometer in
+ 5 % relative humidity balance room and record
temperature daily.
Keep hygrometer in the
balance room and record
humidity daily.
Equipment
Sampler Purchase specification 40CFR. Pt 53.9
Detection Limit Purchase The lower limit of the mass
EDE? rate transfer >j®fci®piti0hle »t^}iPeW, App M. S 7.3 40 CFR, Pt 50, concentration is determined
standard Purchase App M, S 3.1 by the
Analytical balance specification 40 CFR, Pt 50, App M, S 7.5 repeatability of filter tare
Purchase Vol II, MS 2.10.4 weights, assuming the
Mass reference standards specification nominal air sample
Vol II, MS 2.10.4 volume for the sampler.
36
CITY_DEP_WORBY_CANNATA_SUBP_000087319
NYC-WTC_000425895