NYC 9/11 Public Portal Document
Part V - Quality Assurance Project Plan
133-135 Greenwich Street/21-23 Thames Street
Airtek Project 05-0701 - November 4, 2005
Mercury (Gaseous) Using a Direct Read Analyzer
Table 13-lb summarizes the inspection, testing, and maintenance activities associated
with the gaseous mercury analysis using the Ohio Lumex RA915+ mercury analyzer._______________
Table 13-lb. Maintenance, Testing, and Inspection Requirements for the Ohio Lumex RA915 +
Mercury Analyzer
Fault Symptom Possible Cause Corrective Action
Segments of the indicator -Power cable is out of order -Repair the power cable
table on the display and
control unit are not -Display unit cable is out of -Repair the display unit
highlighted when the order
analyzer is switched on.
-Battery is discharged -Charge the battery
The (*) symbol at the -Battery is discharged -Charge the battery
display and control panel is
not dimmed out when the
“Lamp ignition” button is
pressed:
-if the optical switch is in -Operation is possible only -Set the optical switch into positions II
position I with attachments or III
-if the optical switch is in -The single-pass cell is -Remove the single-path cell, check if
position II contaminated, compartment the compartment windows are clean,
windows are contaminated or make sure that there are no foreign
foreign objects are found in the objects inside the compartment
compartment
-Clean the multi-path cell (see
-if the political switch is in -Multi-path cell is appropriate section)
position III contaminated______________
The battery discharge The battery is fully discharged Charge the battery
indicator________________
Fault Symptom Possible Cause Corrective Action
Red (glows for some time
and then goes out when the
aiialyzeiJ^jwitehed oii^
In the TEST mode, relative -Special lamp is not switched -Press the “Lamp ignition” button
deviation R of the measured on -Set the switch of the test cell to the
test number differs from the -The test cell switch is in idle working position
tabulated value by more position -Shake the test cell 2-3 times by the
than 25%. -The test cell is out of order TEST (ON/OFF) switch
-The absorption filter has failed -Replace the filter
86
CITY_DEP_WORBY_CANNATA_SUBP_000072626
NYC-WTC_000424408
OCR can misread numbers and units. Confirm readings against the page image before using them.